E2615-LEL
Rev 24.08.2022
Electrical connections
The terminals on the E2615 series devices are suitable for a wide range of wires with a
cross-section of 0.2...1.5 mm
2
. We recommend to strip the wire end by 8...9 mm and use
wire end sleeves.
To connect the wire, loosen the screw, insert the wire end into the terminal hole and tighten
the screw.
The analog output can be changed between 4-20 mA and 0-10 V type using jumper on the
transmitter’s PCB.
Operation
Turn on the power. The instrument warm-up time takes about 1 minute after switching on
and the final sensor stabilization time to maximum accuracy takes about 30 minutes.
Make sure that the transmitter is properly mounted andthe external devices
connected.Place the lid back and fix it with the screws. The device is ready to use.
Maintenance
Do not perform any maintenance operation with the power on.
Clean the device with a soft damp cloth. Do not use any abrasive cleaning agents. Do not
immerse the device in water or any cleaning media.
11