Description
Out of Range Low Glucose
Appears when your glucose reading is lower than
40 mg/dL.
No glucose reading can be displayed.
Out of Range High Glucose
Appears when your glucose value is higher than
400 mg/dL.
No glucose reading can be displayed.
A
ctions
Confirm the alert on your Eversense CGM
Mobile App.
Measure your glucose manually by
using your blood glucose meter. Always confirm
your glucose value with a blood glucose meter
before making a treatment decision. Once the
sensor glucose value is at or higher than
40 mg/dL, glucose readings will resume on
the display.
Confirm the alert on your Eversense CGM
Mobile App.
Measure your glucose manually by
using your blood glucose meter. Always confirm
your glucose value with a blood glucose meter
before making a treatment decision. Once the
sensor glucose value is at or lower than
400 mg/dL, glucose readings will resume on
the display.