Instruction Manual MM850099 ed.02a
_________________________________________ 18 ________________________________________
7
ELECTRICAL CONNECTIONS (input-output)
Make sure that IN/OUT terminals are connected with the correct polarity.
7.1
Wiring practice
Although the
UniCal mA
portable calibrator is designed to be insensitive to transients or noise, the
following recommendations should be followed to reduce ac pick up in the signal leads and to ensure
good performance.
The input leads should not be run near ac line wiring, transformers and heating elements.
Input/output leads should, if possible, be twisted and shielded with the shield grounded at the end of the
cable.
When shielded cables are used the shield must be connected to the positive terminal.
GENERATOR MODE
(PASSIVE EXTERNAL
CIRCUIT)
Indicator
+
-
P.S.
TRX
+
-
Indicator
+
-
P.S.
TRX
+
-
MEASURING MODE
Indicator
+
-
P.S.
TRX
+
-
Indicator
+
-
P.S.
TRX
+
-
GENERATOR MODE
POWER
(PASSIVE EXTERNAL
CIRCUIT)
Indicator
+
-
P.S.
TRX
+
-
Indicator
+
-
P.S.
TRX
+
-
SIMULATOR MODE
(ACTIVE EXTERNAL
CIRCUIT)
Indicator
+
-
P.S.
TRX
+
-
Indicator
+
-
P.S.
TRX
+
-
POWER SUPPLY MODE
Indicator
+
-
TXR