![EUROIMMUN Medizinische Labordiagnostika AG MERGITE! Instructions For Use Manual Download Page 55](http://html1.mh-extra.com/html/euroimmun-medizinische-labordiagnostika-ag/mergite/mergite_instructions-for-use-manual_2437749055.webp)
6 Operation
Instructions for use
•
MERGITE!
55/84
6.7 Switching the device off
Switching the device off
Turn off the device if you will not use it for an extended time (e.g. at
the end of the work day). In the switching-off process, the liquid
system is automatically rinsed and filled with distilled water in order to
prevent salt deposits.
1) In case of visible contamination, clean the wash basin with a
moist, lint-free cloth. Use a disinfectant for this purpose (e.g.
Meliseptol
®
New Formula).
2) Press the On/Off switch.
o
If the device is already rinsed with distilled water during
monthly maintenance, the device turns off right away.
o
Otherwise you will be prompted to place the wash basin cover
on the wash basin.
3) Place the wash basin cover on the wash basin.
4) Press
START
.
The device is rinsed with distilled water and turned off.
The device has been switched off.
Fig. 38 SHUT DOWN view