TICKET HEADINGS – LOGGING MODE
READING TRIM
Each input can have a ±2.0 °C (3.6 °F) trim value applied to it. This is to
allow for variations in individual thermocouple probe accuracies. Apply
the probes to a known temperature source and adjust the trim values so
that the instrument reading matches the temperature source value. (Use
a temperature source value in the region of the anticipated measurement
values).
For best accuracy the temperature the instrument and temperature
source must be at equilibrium. The source value should be read from a
certified thermometer. The trim value is changed using the drop down
box adjacent to the input name.
The information in the heading fields of the
printed ticket can be changed to suit the user
requirements. These can be changed in the
‘Ticket headings’ section of the Advanced Tab.