background image

3101 Dry-Well

heat/cool source calibrator

User Manual

order code: 271-401

Summary of Contents for 271-401

Page 1: ...3101 Dry Well heat cool source calibrator User Manual order code 271 401...

Page 2: ...2...

Page 3: ...4 safety 4 parts control 5 power lead 5 power switch 5 fan 6 stand 6 dry well 6 temperature controller 6 operation 7 temperature setting 7 probe testing 7 maintenance 8 care of the 3101 Dry Well 8 war...

Page 4: ...pon demand The required temperature may be adjusted in 0 1 C increments across the range of the instrument using the buttons on the front panel These units are ideal for use in a variety of industrial...

Page 5: ...connected to grounded AC power supply 100 240VAC 50 60Hz The unit requires 160 watts of power Mains plug fuse rating is 3 amps The 3101 dry well heat source is programmed and calibrated at the factor...

Page 6: ...storing carefully The inserts MUST be regularly removed and cleaned to ensure they do not seize in the dry well DO NOT introduce any liquids or substance into the dry well or inserts as this may resul...

Page 7: ...required to reach a higher set point temperature It will take longer to reach a lower set point temperature Once the set point temperature is reached a further 10 minutes should be allowed for stabili...

Page 8: ...ire service or repair please contact ETI for assistance There are no user serviceable parts inside and any attempted repair will void any guarantee GUARANTEE This instrument carries a one year guarant...

Page 9: ...r 10 minutes stabilisation 5 minutes cooling times ambient to 0 C after 10 minutes hole dimensions 2 x 13 mm that will accept various inserts inserts each unit is supplied with two inserts of customer...

Page 10: ...ated against reference instrument s calibrated by a UKAS Accredited Calibration Laboratory at 10 C 0 C and 100 C To BS EN 60584 INSTRUMENT RECORD CARD 3101 Dry Well order code 271 401 manufactured by...