Manual Override System

1.

   To use the manual override, remove the access panel labeled “Manual 

Override” on the top of the instrument handle. The manual override 
lever will be exposed.

2.

  Apply force to move the lever forward and backward until it can no 

longer be moved.

ONLY

 to be used if the reverse switch does not return the knife blade.

Reloading the Device

1.

   Push upward (toward the anvil) to unsnap the reload from the reload jaw.

2.

   Prior to reloading the instrument, hold the instrument in a vertical position, 

with anvil and reload jaw completely submerged in sterile solution. Swish 
vigorously and then wipe the inside and outside surfaces to clean any 
unused staples from the instrument. 

3.

    Follow the same steps outlined earlier in the 

“Loading the Device”

 section.

Summary of Contents for ECHELON FLEX Powered

Page 1: ...In Servicing Guide ECHELON FLEX Powered and Powered Plus Staplers...

Page 2: ...se Button Knife Reverse Switch Manual Override Articulation Joint Shaft Articulation Fin and Rotating Knob Reload Alignment Slot Firing Trigger Closing Trigger Red Firing Trigger Lock Knife Blade Indi...

Page 3: ...ining cap by lifting it straight up and discard Loading the Device 3 Ensure that the staple retaining cap is present Insert the reload by sliding it against the bottom of the reload jaw until the stap...

Page 4: ...re on the anvil release button and slowly release the closing trigger Opening and Closing the Jaws Articulation Four Ways to Articulate Ensure that the jaws are in the open position 1 Pull back and ho...

Page 5: ...ng the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation 3 Pull back the red firing trigger lock 4 Pull the exposed trigger to fire m...

Page 6: ...reverse switch forward 2 Battery Remove and re install the battery to check that it is securely installed and that the device has power 3 Reverse See Step 1 4 Override If the knife does not return use...

Page 7: ...if the reverse switch does not return the knife blade Reloading the Device 1 Push upward toward the anvil to unsnap the reload from the reload jaw 2 Prior to reloading the instrument hold the instrume...

Page 8: ...reen D Pro 3 10 bleach 70 isopropyl alcohol prepared according to the manufacturer s instructions Thoroughly scrub areas that contain crevices Thoroughly wipe off detergent with lukewarm tap water DO...

Page 9: ...For complete product details see Instructions for Use 2014 Ethicon US LLC All rights reserved 019927 140814...