Page 2: ...Pump Connections AG Select B1 If using 4 20mA output pump should be connected to J21...
Page 3: ...Manual AG Select B1 J21 mA 1 output J20 mA 2 output...
Page 4: ...Pump Connections AG Select B1 If using 4 20mA output pump should be connected to J21 as shown above...
Page 5: ...Calibration pH AG Select B1 Please ensure you have to hand pH 7 and pH 4 buffer solutions and check use by dates...
Page 6: ...on mix slowly and then wait until display stabilises after around 1 minute Use the PLUS and MINUS buttons to trim the reading to pH 7 00 Press OK button to confirm and to go to next calibration point...
Page 7: ...Calibration mV AG Select B1 Please ensure you have to hand 650mV or 475mV buffer solutions and check use by dates...
Page 8: ...til display stabilises after around 1 minute Use the PLUS and MINUS buttons to trim the reading to mV 650 Press OK button to confirm and to go to next calibration point for ZERO mV To simulate 0 mV us...
Page 9: ...hin 15 seconds 4 After screen has changed hold down the MINUS and PLUS buttons simultaneously 5 rESEt should be shown on screen if you do not see SET DEFAULTS start again taking extra care on step 4 6...