ETAS
Hardware Description
ES620.1 - User Guide
13
The modules are configured using the INCA’s graphical user interface (GUI).
INCA also provides you the functionality for data acquisition and data evalua
-
tion.
3.2
ES620.1 Thermo Module
3.2.1
Overview
The ES620.1 is designed to measure temperatures using thermocouples. It
provides you with 16 channels for temperature measurement with thermocou
-
ples.
Fig. 3
-
1
ES620.1 Thermo Module
The thermocouple channels have a resolution of 21 bits.
You can use B-, E-, J-, K-, N-, R-, S- and T-type thermocouples. The maximum
measuring range is
-
210 °C to +1,820 °C. Each thermocouple channel is fitted
with a cold junction compensation directly in the connector. This ensures accu
-
racy even when the housing is not heated uniformly. The thermocouples are
galvanically isolated from each other and from case ground.
NOTE
The ES620.1 only switches from the power-saving standby mode to opera
-
tion mode when it receives link pulses via the Ethernet interface. To ensure
that a connected PC sends link pulses, its Ethernet adapter must be properly
configured. A data transfer via the Ethernet interface automatically switches
the ES620.1 into operation mode.
TH1
ON
TX
ER
ES62
0.1
TH9
TH2
TH10
TH3
TH11
TH4
TH12
TH5
TH13
TH6
TH14
TH7
TH15
TH8
thermocouple
0463
318