ETAS
Getting Started
ES413.1 - User’s Guide
35
5.1.3
Guarantee of Features as Defined by IP65
If installing the ES400 modules where water or other liquids could collect,
install the modules so that the (black) pressure balance element on the back of
the modules is not pointing upwards or that liquids can drain away.
Fig. 5
-
1
Position of the Pressure Balance Element
Standing liquids or liquids which do not flow away from the pressure balance
element can permanently damage the membrane. The module then loses the
features defined by IP65.
5.2
Assembly
CAUTION
Loss of Features as defined by IP65!
Water standing at the pressure balance element damages the membrane!
Observe which way the module is pointing when installing vertically!
TIP
ES413.1 and ES421.1-K modules you can mechanically interlink or connect
onto other components.
You need additional parts (srews or threaded rods with nuts, cable fasteners
etc.) to assemble the modules.
TIP
ES413.1 and ES421.1-K modules you can only mechanically interlink with
other ES400 modules (housings with integrated assembly elements) at the
openings for cable fasteners.
DAE