ES1652.2 Carrier Board for Signal Conditioning Circuits - User’s Guide
15
ETAS
Technical Data
4
Technical Data
Voltages, Currents and Power Consumption
Environmental Conditions
Physical Dimensions
4.1
Fulfilled Standards and Norms
The ES1652.2 Carrier Board for Signal Conditioning Circuits complies with the
following standards and norms:
Maximum voltage at inputs and outputs
60 V
Max. current per individual contact of the input/output connectors
1 A
Maximum sum current over input/output connectors
5 A
Maximum permissible power consumption per DC/DC converter
3 W
Maximum permissible power consumption from VMEbus backplane 10 W
Operating temperature
+5 °C to +50 °C (+41 °F to +122 °F)
Relative humidity
0 to 95% (non-condensing)
Storage temperature
+5 °C to +85 °C (+41 °F to +185 °F)
Relative humidity
0 to 95% (non-condensing)
Height
3 U
Width
4 HP
Norm
Test
EN 61326-1
Electrical equipment for measurement, control and
laboratory use – EMC requirements
EN 61000-6-2
Immunity (industrial environments)
EN 61000-6-3
Emission standard (residential, commercial and
light-industrial environments)
WARNING!
The device confirms only with radio disturbances class A (only for industrial
environment). This can cause radio disturbances in residential, commercial
and light industry environment. In such a case the operator has to consider
additional measures for shielding to avoid disturbances.
Summary of Contents for ES1652.2
Page 1: ...ES1652 2 Carrier Board for Signal Conditioning Circuits User s Guide...
Page 4: ...ES1652 2 Carrier Board for Signal Conditioning Circuits User s Guide 4 Contents ETAS...
Page 16: ...ES1652 2 Carrier Board for Signal Conditioning Circuits User s Guide 16 Technical Data ETAS...
Page 20: ...ES1652 2 Carrier Board for Signal Conditioning Circuits User s Guide 20 Index ETAS...