Intended Use and User
The scope was developed to monitor biological samples in the laboratory. For this, the scope
is attached directly to the respective vessel or sample and carried along by the complete
process. In addition to use in the laboratory, the scope can also be used for transport
monitoring, as long as the safe operating conditions are not
surpassed or under-passed
.
Warning!
Always follow the stated safe conditions of use!
●
The operating temperature is -20 °C to + 60 °C with non-condensing humidity up to 100 %
and an ambient pressure of 0.8
bar
to 1.2 bar.
●
Scopes should be used as a stationary or mobile measuring unit.
●
Scopes must be charged at room temperature (but in any case between 0 °C and 40 °C).
●
Only use
essent
im
approved cables and extension in conjunction with the scope.
●
Avoid mechanical stress on the scope, damage to the housing may damage the
lithium-polymer battery. This can lead to leakage of liquids from the battery or short
circuits and temperature development.
●
For transport monitoring, the transport destination of the respective transport service
provider with regard to lithium-containing batteries must be considered. For more
information contact your transport service or
essent
im
GmbH.
The typical user is e.g. a trained laboratory assistant / assistant / technician / student. Every
user must read and understand the safety and the user manual.