Checks and Maintenance
80
Platinous User’s Manual
7.3.5 Test Area Cleaning
Dirt and foreign matter inside the test area can throw test results off. Clean the test area before every test.
Procedure:
1. Open the chamber door.
2. Wipe walls and parts clean with a soft cloth.
3. Close the door.
7.3.6 Electrical Compartment and Water Circuit Compartment
Cleaning
Dust buildup inside the electric parts compartment and the water circuit compartment can cause problems.
Clean inside both compartments once a month.
Procedure:
1. Check that the main power breaker is in the OFF position.
2. Open the electric parts compartment and water circuit compartment doors.
3. Clean dirt from both compartments with a vacuum or by other means.
4. Close the compartment doors.
7.3.7 Take-Down Before Long Periods of Disuse
Before long periods of disuse, do the following. Failure to do so can affect testing and shorten equipment
service-life.
Drain the wick pan and humidifying tray of water by opening the drain valve. (Not necessary with
temperature-only chambers).
FIGURE 7.48.
D
RAIN
VALVE