Installation
36
Platinous User’s Manual
5.1 Installation Site Check
This section describes installation site and space requirements.
Installation Site
Install the chamber in a place which satisfies the following conditions:
•
On a flat, level floor which is strong enough to bear the weight of the equipment.
•
Where subject to only minimal mechanical vibrations.
•
Where not exposed to direct sunlight but which is well-ventilated.
•
Where ambient temperature is +10 to 25°C and free of sharp temperature fluctua-
tions. (0.1°C/min. or 3°C/30 mins.)
•
In a dust-free room.
•
Where not exceptionally wet or humid.
•
Away from flammables and explosives.
•
Where not exposed to combustible or corrosive gases.
•
Near power, water, and drainage utilities.
•
Not directly underneath or near fire alarms.
Note:
Chamber working temperature range is 0-35°C. Using the chamber outside of this
range will lead to equipment problems. Performance is only guaranteed when in
a room 10 to 25°C.
Installation Space
The chamber requires operational and maintenance space on the left and right sides. It must also be a
safe distance from objects on the front side so that the door can be opened.
Summary of Contents for Platinous ESL Series
Page 1: ...PLATINOUS CHAMBER SERIES USER S MANUAL...
Page 2: ......
Page 6: ...iv Platinous User s Manual...
Page 22: ...Precautions in Usage 16 Platinous User s Manual...
Page 31: ...Platinous User s Manual 25 Names and Functions of Parts...
Page 40: ...Names and Functions of Parts 34 Platinous User s Manual...
Page 58: ...Installation 52 Platinous User s Manual...
Page 80: ...Operation 74 Platinous User s Manual...
Page 106: ...Troubleshooting 100 Platinous User s Manual...