14
Example:
After specifying the S-LO value, you can do the same for the S-HI value.
6.2.3
Calibration Method Determination
Your device has been calibrated during production. If you want, you can do this calibration
according to the sensor. You need to select the calibration method from this menu.
6.2.3.1
Factory calibration (default)
The default calibration on your device. The multipoint calibration or the two points you
make your device does not disturb the factory calibration.
(Press and hold the PRG
key while the device is in
operating mode)
Factory calibration (Default)
2 point calibration
Segmented calibration
.
.
.
.
.
The decimal point
starts flashing. Use
the UP and DOWN
keys to select the
decimal point.
Press and hold the UP and DOWN
keys to move to the next step. To
enter a negative (-) value, press
the DOWN key to come to one of
the non-digitized steps.
The value in the
second line starts
flashing. Enter the
desired value using
the UP and DOWN
keys.
.
.
.
.
The new S-LO value you confirmed with the PRG key has been
set. Press UP button to move to S-HI.