ESD SYSTEMS 41290 Operation And Maintenance Download Page 1

Surface Resistance Test Kit 
Operation and Maintenance

Tech Brief

PS-2111

September 2007 Page 1 of 7

Made in America

Description

The Surface Resistance Test Kit is a
portable, accurate, and versatile
instrument designed to measure
resistance between two points (RTT),
surface to ground (RTG), and surface
resistivity in accordance with ESD
Association Standard S4.1 including:
• Resistance measuring accuracy ± 10%

(E11 and greater ±20%)

• Resistance range <1.0 x 10E3 ohms

to >10E12 ohms

• Open circuit voltages of 10 and 100

volts ± 5%

• Electrification period of 15 seconds
• Electrodes (two) 5 pounds ± 2 oz with

50-70 durometer conductive pads

In addition, the Meter measures ambient
temperature and relative humidity.

The Surface Resistance Test Kit (or its
Meter) is referenced and designed to be
used to make measurements in
accordance with the test methods in:
• Worksurfaces - ANSI/ESD S4.1

Worksurfaces Compliance
Verification - ESD TR53 - Resistance
Measurements

• Floors - ANSI/ESD S7.1- Resistive

Characterization of Materials Floor
Materials

• Footwear - ESD S9.1 - Footwear -

Resistive Characterization

• Garments - ESD STM2.1 Garments
• Seating - ESD STM12.1- Seating -

Resistive Measurement

• Floor/Footwear - ESD STM97.1 -

Floor Materials and Footwear-
Resistance Measurement in
Combination with a Person

• Workstations - ESD-ADV53.1 - ESD

Protective Workstations

Versatility:

Measures RTT, RTG or Resistivity
(requires optional Resistivity
Attachment, item 41292)
• 

Quick Checks -

Surface resistance

exponent number illuminates
immediately i.e. 8 = 10E8 ohms or
100,000,000 ohms

Periodic Checks of Installed
Products -

Surface resistance

exponent number illuminates
immediately and displays
Temperature and Relative Humidity
during 15 second electrification
period making numerous measure-
ments and calculations to then display
mantissa i.e. if LED displays 8 and
LCD displaying 7.14 as mantissa =
7.14 x 10E8 ohms or 714,000,000
ohms

Test Lab Evaluation of Product -

Display same as Periodic Check
above

The Surface Resistance Test Kit is NIST
calibrated and items available are:

Inspection

Remove the meter from the carton and
inspect for damage. Each Surface
Resistance Test Kit should include the
following:
1  Protective carrying case
1 Meter
2 Test 

leads

2 5-pound 

electrodes

2 AA alkaline 

batteries

Properly store the Meter and component
assemblies when not in use. 

Figure 1. Item 41290 Surface Resistance Test Kit

© 2007 DESCO INDUSTRIES INC.

Employee Owned

Model

Description

41290

Surface Resistance Test Kit

41291

Meter & Attachment 

41292

Resistivity Attachment

Summary of Contents for 41290

Page 1: ...loor Footwear ESD STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person Workstations ESD ADV53 1 ESD Protective Workstations Versatility Measures RTT RTG or Resistivity requires optional Resistivity Attachment item 41292 Quick Checks Surface resistance exponent number illuminates immediately i e 8 10E8 ohms or 100 000 000 ohms Periodic Checks of Installed Product...

Page 2: ...ing displayed is identified by the illuminated function LED The surface resistance exponent or power of number is immediately illuminated and remains illuminated measuring the range the surface resistance in ohms Reference Literature In addition to those noted above ANSI ESD S20 20 Development of ESD Control Program ESD ADV1 0 Glossary of Terms ANSI ESD S6 1 Grounding These documents can be obtain...

Page 3: ...urfaces use ESD Systems com Reztore Antistatic Surface and Mat Cleaner item 10630 or other ESD cleaner not containing silicone Be sure the surface is dry before testing Periodic maintenance The area surrounding the cable jacks at the top end of the meter should be wiped with a clean cloth moistened with alcohol to remove skin oils that will accumulate and affect the accuracy at high resistances Th...

Page 4: ...ulative silicone i e Reztore Antistatic Surface and Mat Cleaner Reporting and Using Test Results Different standards have different requirements use requirements as specified in user s ESD control plan Examples are Per ESD 4 1 worksurfaces Report RTG maximum and minimum values measured for resistance to ground RTT maximum and minimum values measured for point to point resistance in ohms Per ESD 7 ...

Page 5: ...ing Diodes LED s illuminate exponent numbers and function being performed Accuracy Resistance measurements within 10 E12 and greater 20 complies with ESD S4 1 Open circuit voltages of 10 volts 5 and 100 volts 5 exceeds requirements of ESD S4 1 Meter Weight 15 5 oz Dimensions 8 L x 4 3 W x 1 6 H Test Switch Press the red test button for 15 20 second electrification period Meter will immediately dis...

Page 6: ...o AUTO switch up Testing each decade starting from 1 0 x 10E12 and down never start from 1 0 x 10E3 Temperature Fahrenheit 75 F 6 6 Celsius 23 8 C 10 2 Relativity Humidity 10 Digits A 1 0 x 10E12 20 LED 12 Yellow Mantissa 1 20 0 LED 12 Yellow Mantissa 1 00 20 LED 11 Yellow Mantissa 8 00 B 1 0 x 10E11 20 LED 11 Yellow Mantissa 1 20 0 LED 11 Yellow Mantissa 1 00 20 LED 10 Green Mantissa 8 00 C 1 0 x...

Page 7: ...390 for a Return Material Authorization RMA and proper shipping instructions and address Desco will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The expr...