64
Table 6-11. The syntax of returned character string
Function
Range
Return Value
500mV
±
500.00E-3
5V
±
5.0000E+0
50V
±
50.000E+0
500V
±
500.00E+0
DC Voltage
1000V
±
1000.0E+0
500mV +500.00E-3
5V +5.0000E+0
50V +50.000E+0
500V +500.00E+0
AC / AC+DC
Voltage
750V +0750.0E+0
500
μ
A
±
500.00E-6
5mA
±
5.0000E-3
50mA
±
50.000E-3
500mA
±
500.00E-3
5A
±
5.0000E+0
DC / AC /
AC+DC
Current
10A
±
10.000E+0
500
Ω
+500.00E+0
5K
Ω
+5.0000E+3
50K
Ω
+50.000E+3
500K
Ω
+500.00E+3
5M
Ω
+5.0000E+6
Resistance
50M
Ω
+50.000E+6
500Hz +500.00E+0
5KHz +5.0000E+3
50KHz +50.000E+3
Frequency
500.00E+3
Note: The return string of
±
9E+9 means overload condition.
LLO command
LLO
command is used to lock local key operation. The K13 will not been used after
this command.
GTL command
GTL command is used to exit remote state and go to local. The K13 will be available
then.
BON command
BON command is caused the meter to sound one tone.