5-4 Recommended Test Equipment
The test equipment requirements listed in Table 5-1 or equivalents are
required to perform the calibration and performance verification test
procedures. Alternative equipment may be used as long as the accuracy is at
least as good as those listed.
Table 5-1 Standard Equipment Requirements
Standard
Source
Operating
Range
Accuracy
Required
Recommended
Equipment
DC Voltage
Calibrator
Range, 0 to 1000VDC
≤
± 0.002%
Fluke 5520A or
equivalent
AC Voltage
Calibrator
Range, 0 to 750V, 1kHz
≤
± 0.03%
Fluke 5520A or
equivalent
10mA to 100mA
≤
± 0.01%
DC Current
Calibrator
1A to 10A
≤
± 0.03%
Fluke 5520A or
equivalent
10mA to 1000mA, 1kHz
≤
± 0.1%
AC Current
Calibrator
1A to 10A, 1kHz
≤
± 0.2 %
Fluke5520A or
equivalent
450
Ω
, 4.5k
Ω
, 45k
Ω
,
450k
Ω
, 4.5M
Ω
≤
± 0.01%
Fluke 5520A or
equivalent
Resistance
Calibrator
30M
Ω
≤
± 0.05%
Fluke 5520A or
equivalent
Audio Level
Generator
2V/4500Hz
≤
± 0.005%
Fluke 5520A or
equivalent