Operation Example:
Using MIN.MAX Modifier for Dynamic Recording
“MIN MAX”
modifier enables the meter to store the minimum and
the maximum input signals measured and elapsed time the value recorded
since the
“MIN MAX”
modifier was selected. The definitions of
“MIN”
and
“MAX”
are defined as follows:
MIN: Minimum value of calculation results for measured signal
MAX: Maximum value of calculation results for measured signal
If
“MAX”
modifier is selected, the display indicates the latest maximum
reading and elapsed time accordingly until next measurement reading exceeds
the previous recorded reading.
If
“MIN”
modifier is selected, the display indicates the latest minimum reading
and elapsed time accordingly until next measurement reading drops below the
previous recorded reading.
If
“MIN MAX”
mode is selected, the display indicates the actual value of input
signal. Selecting this modifier in auto range, it will record the value of MAX,
MIN for different ranges.
The elapsed time is recorded since the modifier has been selected.
The
elapsed time is shown on the secondary display with
“HH.MM.SS”.
The HH is
0~19 hours, MM is 0~59 minutes and SS is 0~59 seconds.
Note: If the beep mode is set to “ON”, the beeper will emit a single tone when
an effective maximum or minimum value is recorded.