96
AC Current (True RMS, AC Coupling)
Resolution, Full Scale Reading and Burden Voltage
Accuracy (1 year)
(5)
Range
Resolu-
tion
Full Scale
Reading
30 to
50 Hz
50 to
2k Hz
2k to
5k Hz
5k to
20k Hz
Burden Voltage
(1)
&Shunt Resistor
500
µ
A
5mA
50mA
500mA
10nA
100nA
1
µ
A
10
µ
A
510.00
5.1000
51.000
510.00
1.5% + 50
1.5% + 40
1.5% + 40
1.5% + 40
0.5% + 20
0.5% + 20
0.5% + 20
0.5% + 20
1.5% + 50
1.5% + 40
1.5% + 40
1.5% + 40
3% + 75
(4)
3% + 60
3% + 60
3% + 60
<0.06V/ 100
Ω
<0.6V / 100
Ω
<0.08V / 1
Ω
<0.8V / 1
Ω
5A
10A
100
µ
A
1mA
5.1000
20.000
(2)
2% + 40
(3)
2% + 40
(3)
0.5% + 30
(<1kHz)
Not Specified
<0.3V / 0.01
Ω
<0.6V / 0.01
Ω
(1)
Typical at full scale reading and voltage across the input terminals
(2)
In 10A range, >10~20Aac is readable for 20 seconds maximum with audio warning.
(3)
Input Current < 3 Arms.
(4)
Input Current > 35
µ
Arms.
(5)
Accuracy specified at input >5% of range and >1A for 10A range except other specified.
•
Measurement method: True RMS
•
Maximum Crest Factor: 3.0 at full scale
•
Response Time: Approximately 1.5 seconds when the displayed reading reaches 99.9%
ac rms value of the tested input signal at the same range.
AC Voltage (True RMS, AC+DC Coupling)
Resolution, Full Scale Reading and Accuracy
Accuracy
(1 year)
(2)
Range Resolution
Full Scale
Reading
50 to 10k Hz
10k to 30k Hz
30k to 100k Hz
500mV
5V
50V
500V
750V
10
µ
V
100
µ
V
1mV
10mV
100mV
510.00
5.1000
51.000
510.00
1000.0
(1)
0.5% + 50
0.5% + 25
0.5% + 25
0.5% + 25
0.5% + 25
(4)
2% + 70
1% + 30
1% + 30
1% + 30
(3)
1% + 30
(3)
3% + 130
3% + 60
3% + 60
3% + 60
(3)
Not Specified
(1)
In 750V range, 1000.0V is readable with audio warning.
(2)
Accuracy specified at input >5% of Range.
(3)
Input Voltage < 200V rms.
(4)
For 5k ~10 k Hz, the accuracy is 0.7%+25.
•
Measurement method: True RMS
•
Maximum Crest Factor: 3.0 at full scale
•
Maximum input voltage: 1000V rms, 1400V peak ac
2x107 V-Hz product on any range, normal mode input
1x106 V-Hz product on any range, common mode input
•
Input Impedance: 1M
Ω
in parallel with capacitance <100pF
•
Response Time: Approximately 1.5 seconds when the displayed reading reaches 99.9%
ac rms value of the tested input signal at the same range.