TL family`s multiroom IVF incubators User Manual Rev. 1.0
38
The default status for O
2
control is "OFF".
The O
2
flow rate is shown when the setpoint value is set. The flow rate cannot be adjusted
because this is the amount of O
2
gas put into the system while regulating. The volume is
shown in liters per hour. It usually will fluctuate along with the CO
2
regulation.
The O
2
pressure value is shown in bar. The external pressure must be between 0.4
–
0.6
bar (5.80
–
8.70 PSI) at all times. It cannot be adjusted on the multiroom IVF incubator; it
must be done on the external gas regulator.
13.5.4 Temperature Setpoint sub-menu
Choose temperature setpoint settings according to the desired working conditions.
If "Multi Temp SP" is "ON", it means that temperature values in compartments are
individual. If "Multi Temp SP" is "OFF", it means that temperature values in compartments
are combined.
Figure 13.27
Multi-temperature setpoint view in the MIRI
®
TL6 and MIRI
®
TL12 multiroom IVF
incubator
When the desired option is set, press the "EXIT" button. The option will be saved
automatically.
When the MULTI temperature setpoint is “OFF”, all compartments' temperature
setpoint is set according to T1 by default setting. If the MULTI temperature setpoint
is “OFF”, changing any compartment's temperature setpoint value will apply to all
remaining compartments. If the MULTI temperature setpoint is “ON”, each
compartment has different setpoint values. When returning from “ON” mode to
“OFF”, all setpoints are automatically set to the T1 area value.
It is recommended to keep the MULTI temperature settings to “OFF” if all
compartments run at the same temperature. It will be easier to make adjustments
to the setpoint as the adjustment will only have to be done once instead of twelve
times (i.e., for individual compartments).