MIRI
®
TL family`s multiroom IVF incubators User Manual Rev. 1.0
112
38.4 Bring the following to the installation site
•
"Installation report" form.
•
Service manual for the MIRI
®
TL6 or MIRI
®
TL12 multiroom IVF incubator.
•
Updated service tool kit.
•
Memory stick with the latest released firmware & software.
•
High precision thermometer with a resolution not less than 0.1 °C.
•
Calibrated gas analyzer with precision at least 0.1% for CO
2
and O
2
and the
possibility of returning gas samples to the incubator.
•
Extension cable for USB connection.
38.5 Installation procedure at the site
1.
Follow the guidelines in the safety instructions and warnings section ("2 Safety
warning" section).
2.
Connect the power cable to the UPS.
3.
Connect the power cable to the MIRI
®
TL6 or MIRI
®
TL12 multiroom IVF incuba-
tor.
4.
Connect the gas lines.
5.
Set gas pressure on the external gas regulator at 0.4 – 0.6 bar (5.80 – 8.70 PSI).
6.
Switch on the MIRI
®
TL6 or MIRI
®
TL12 multiroom IVF incubator on the back.
7.
Observe for standard functionality.
8.
Let the unit warm up and stabilize for 30 min.
9.
Follow the guidelines in the “33 The Validation guide” section.
10.
Complete user training and finish reading instructions.
11.
After a burn-in phase of 24-hours, the unit is ready for use IF the testing is suc-
cessful.
38.6 User training
1.
Mains switch on/off.
2.
Explain the MIRI
®
TL6 or MIRI
®
TL12 multiroom IVF incubators essential function
and incubation with a multi-room facility to store the samples.
3.
Explain temperature control in the MIRI
®
TL6 or MIRI
®
TL12
multiroom IVF incu-
bator (direct heat transfer with heated lids).
4.
Gas regulation on/off.
5.
Setpoint for temperature, CO
2
and O
2
.
6.
Explain how N
2
is used to suppress the O
2
level.
7.
Alarm turn off procedure (temperature, CO
2
, O
2
) and revert times.
8.
Emergency procedures (can be found in the "29 Emergency Procedures" section).
9.
Explain how to clean the device.
10.
External measurement and calibration of temperature.
11.
External measurement and calibration of gas concentration.