MIRI
®
II-12 Multiroom IVF incubators User Manual Rev. 4.0
81
37.15 Firmware update
If Esco Medical has released a newer version of the firmware, this should be installed on the
MIRI
®
II-12 multiroom IVF incubator during the yearly scheduled service.
Please refer to the service manual for instructions on how to update the firmware.
38 The Installation guide
This section describes when and how to install the MIRI
®
II-12 multiroom IVF incubator in
the IVF clinic.
38.1 Responsibilities
All technicians or embryologists installing the MIRI
®
II-12 multiroom IVF incubator must
identify problems and perform any necessary calibrations, adjustments and maintenance.
Installation personnel performing MEA (Mouse Embryo Assay) must be thoroughly familiar
with the MEA and all functions of the instrument, calibration and testing procedures, and
instruments used in the instrument's testing. MEA test is a supplemental installation test
and is not mandatory.
All individuals who will perform installation, repair and/or maintenance of the instrument
must be trained by Esco Medical or at a qualified training center. Experienced service
technicians or embryologists conduct training to ensure that the installation personnel
clearly understand the instrument's functions, performance, testing, and maintenance.
Installation personnel must be updated regarding alterations or additions to this document
and the "Installation report" form.
38.2 Before installation
2 – 3 weeks before the installation due, the user/owner at the clinic is contacted via e-mail
to plan the exact time to perform the installation. When a convenient time has been
determined, travel and accommodation arrangements can be made.
The released the MIRI
®
II-12 multiroom IVF incubator must be sent 1 – 3 weeks before
installation, depending on the clinic location. Check with shippers about local customs
regulations and delays that could arise from that.
The clinic must be informed about the site requirements before installation and should have
signed the customer requirement checklist:
1.
The lab must have an idle sturdy and stable lab bench for standing operation.