MIRI
®
II-12 Multiroom IVF incubators User Manual Rev. 4.0
76
After 1-year, the chemical process in the sensor core stops and the sensor must be replaced.
Therefore, it is essential to replace this sensor
WITHIN year from the date it was
unpacked and installed.
Oxygen sensors must be replaced at least once every year from the date they were
installed in the unit, irrespective of the incubator being used or not.
In the MIRI
®
II-12 multiroom IVF incubator "Installation report" form, the User will see
when this sensor was installed. This date must be used to calculate the date for the following
O
2
sensor change.
Please follow these safety precautions when changing sensor:
•
Always use an original O
2
sensor (contact Esco Medical or your local distributor for
more details or ordering).
•
Change the O
2
sensor within 1 year from the date of the previous sensor installation.
•
Failure to change the oxygen sensor on time will result in low/no regulation of O
2
concentration.
•
Warranty void if wrong/ non-original sensor is used.
Please refer to the service manual for replacement instructions.
37.5 CO
2
sensor
The CO
2
regulation uses the CO
2
sensor to keep the gas concentration at the chambers'
desired level.
This sensor's lifetime is more than 6 years, but for safety reasons, Esco Medical recommends
the sensor to be replaced once every 4-years.
Please follow these safety precautions when changing the sensor:
•
Always use an original CO
2
sensor (contact Esco Medical or your local distributor for
more details or ordering).
•
Change the CO
2
sensor within 4 years from the date of installation.
•
Failure to change the CO
2
sensor on time can result in low/no CO
2
gas concentration
regulation.
•
Warranty void if wrong/ non-original sensor is used.
Please refer to the service manual for replacement instructions.