9 Maintenance 

Esco  Medical  AVT-1

®

  is  designed  to  be  easy  to  use  and  requires  almost  no  mainte-

nance at all.  
 
Over time, adjustments may become necessary.    
 
The rubber paste should be changed every 2 years, the springs 

– every 6 years. 

 

 Warranty  is  considered  to  be  void  if  service  and  maintenance  procedures 

are not followed. 
 

 Warranty  is  considered  to  be  void  if  service  and  maintenance  procedures 

are done not by a trained and authorized personnel.   

10 User troubleshooting 

Table 10.1 

Heating system 

Vibrations 

Cause 

Action 

 

The  float  is  touching  the  table 
plate  and  vibrations  are  trans-
ferred directly to the microscope.  

The  floater  should  be  in  the 
centre  position  of  the  hole. 
Air  should  be  all  around  the 
floater.   

 

Loose tightening bolts. 

Tighten four tightening bolts. 

The  spring  system  is  not 
painted  and  has  to  be 
completely  grease  free  to 
function.   
The surface is made to be 
rough intentionally.  

Grease  or  oil  on  the  surface  of 
contact points. 

Clean  the  area  with  70%  al-
cohol.   

Tight  cables  of  micro-
manipulators  or  any  other 
equipment that are placed 
on the table top will result 
in  vibrations  being  trans-
mitted  directly  to  the  mi-
croscope. 

Cables routed incorrectly.  

Tie cables so that they would 
not touch the table and would 
hang  loose  between  the  at-
tachment points. 

 

Microscope  is  placed  unevenly 
on the floater.  

The weight of the microscope 
has  to  be  distributed  evenly 
on the floater. Since the back 
part  of  a  microscope  is  the 
heaviest,  it  should  be  moved 
more to the front of the table 
plate.  

Summary of Contents for AVT-1

Page 1: ...Rev 1 8 Rx only Anti vibration Table AVT 1...

Page 2: ...s the right to make periodic minor design changes without obligation to notify any person or entity of such change Caution Federal law restricts this device to sale by or on the order of a licensed he...

Page 3: ...se note that only the serialized products products labelled with a distinct serial number and accessories are eligible for partial refund and or credit Non serialized parts and accessory items cables...

Page 4: ...other than Esco Medical IN NO EVENT SHALL ESCO MEDICAL LTD BE HELD LIABLE FOR CONSE QUENTIAL DAMAGES No warranty shall apply when damage is caused by any of the following Power failure surges or spike...

Page 5: ...original warranty covering your product becomes void when the tamper resistant Quality Seal is removed or broken without proper factory authorization In all cases breaking the tamper resistant Qualit...

Page 6: ...AVT 1 User Manual Rev 1 8 6 COMPONENTS UNAUTHORIZED REMOVAL OF THE INSTRUMENT COVER SHALL VOID THIS AND ANY OTHER EXPRESSED OR IMPLIED WARRANTIES...

Page 7: ...product 8 4 Accessories supplied 9 5 Safety symbols and labels 9 6 Important safety instructions and warnings 10 6 1 Before installation 10 6 2 During installation 10 6 3 Post installation 10 7 Gettin...

Page 8: ...nance to this equipment must be un dertaken by a qualified service personnel If the equipment must be relocated make sure it is fixed properly on a support stand or base and move it on a flat surface...

Page 9: ...s the microscope area The degree of improvement varies between instruments and locations depending on the amplitude and frequency of floor vibrations This product does not carry a CE mark Only those p...

Page 10: ...carpet or similar surfaces 3 This unit is intended for indoor purpose only 4 Lift the device carefully All AVT 1 parts are heavy and may require additional people to move them around 5 Float elements...

Page 11: ...le plate Figure 7 1 Leveling table legs Put on gloves Install the springs into the holes at the front and the back of the table A 3rd spring placed in the centre may be supplied for the use of a very...

Page 12: ...tops should be facing up Leave equal distance from the hole on all sides of the table plate Then place the top of the floating element stainless steel on the rubber parts again leaving equal distance...

Page 13: ...ween the attachment point on the microscope and the other side This can be achieved by strip ping the cable to the points and making sure that the cable forms an arch between the points If the cable i...

Page 14: ...be in the centre position of the hole Air should be all around the floater Loose tightening bolts Tighten four tightening bolts The spring system is not painted and has to be completely grease free t...

Page 15: ...ight 70 kg Material Powder painted mild steel Stainless steel Glass Float Size 540 340 Power consumption 115V 60Hz OR 230V 50Hz Recommended load weight 26 32 kg Damping coefficient 6 Hz 0 1 Amplitude...

Reviews: