16 /
20
Chapter 5
quadrature
bandwidth: gain 1 up to 0.43xfc, with fc
from 156.25 to 4.882 KHz
dark band: from 0.5xfc
attenuation 90 dB
transfer to DSP of two pairs of raw data (real
and imaginary part)
Synthesizer:
digital, with frequency, amplitude and phase
modulation and resolution: frequency
9.3 MHz, amplitude 4096 levels, phase 1’19"
stability: <1 ppm in operating temperature
range
d
8 x 10
-8
in 15 minutes (0.75 degrees
centigrade maximum variation)
variable transmission gain: 16384 levels
RF transmission coil:
linear, solenoidal. input impedance 50 Ohm,
maximum power 1500 Wpep
RF transmission power:
power aplifier up to 1500 W pep RF amplifier
bandwidth 12.7 - 13.3 Mhz at 1 dBpp
Patient positioning
Patient seat:
maximum admissible weight = 150 kg
fixed backrest height and inclination
leg rest
washable covering material
manual positioning
Foot rest device suitable also for cinematic examinations of the knee;
pedal switch for scan start-up.
Comfortable rest for non-examined limb.
Keyboard and display on magnet (optional):
pushbuttons for real-time
management; TFT 3.5" display, 240x320, 64
grayscale levels for the display of real-time
images, to simplify patient positioning
Wrist locking/movement device (optional).
Installation area conditions
Type of installation:
permanent or mobile
Overall dimensions:
2,8 mx 3,7 m
2,5 mx 3,2 m without console
2,3 m x 4,5 m (mobile installation)
Environmental operating conditions:
temperature: 20÷26 °C with variations of
Summary of Contents for O-scan eXP
Page 1: ......
Page 2: ......
Page 3: ...O scan User Manual 350004900 Rev 07 Esaote S p A Genoa...
Page 9: ...350004900 Rev 07 7 10...
Page 10: ...8 10...
Page 12: ...10 10...
Page 20: ...8 8...
Page 32: ...8 8 Chapter 2...
Page 51: ...350004900 Rev 07 19 48 Instructions for Safe Use fig 3 1 Top view...
Page 52: ...20 48 Chapter 3 fig 3 2 Side view...
Page 80: ...48 48 Chapter 3...
Page 102: ...22 22 Chapter 4...
Page 105: ...350004900 Rev 07 3 20 Technical Description fig 5 1 Equipment components...
Page 122: ...20 20 Chapter 5...
Page 142: ...20 20 Chapter 6...
Page 161: ...350004900 Rev 07 5 10 Soft cushion Set fig 9 6 Use of lumbar support cushion...
Page 184: ...10 10 Chapter 11...
Page 232: ...4 4 Chapter 17...
Page 236: ...4 4 Chapter 18...