M A I N T E N A N C E
MyLab
- G E T T I N G S T A R T E D
8 - 2
NOTE
Frequency of cleaning can change depending on environment cleanliness.
NOTE
Periodic maintenance operations that require the access to the device can
be performed only by trained personnel: contact your local Esaote
representative for further information on required periodic inspections.
Only trained persons are allowed to perform the safety inspection mentioned
above.
Disconnect
MyLab
from the power outlet before checking it.
Contact Esaote personnel for any problem found during inspection.
PC
Refer to “Probes and Consumables” manual for periodic inspections for
probes and cleaning instructions.
Veterinary Applications
WARNING
Due to the specificity of the veterinary working place, Esaote recommends
that the ultrasound devices used in veterinary applications should be
thoroughly cleaned at least once every 6 months. Please contact your
Esaote distributor or dealer for further help in this matter.
checking entire length of electrical and
power cables for cuts or damages
every month
user
checking the device integrity,
functionality and cleaning (including
internal components)
every year
Esaote personnel
or authorized
personnel
Electrical Safety Tests
every two years
Esaote personnel
or authorized
personnel
Maintenance
Minimum frequency
Performed by
Summary of Contents for MyLabX75VET
Page 1: ...Rev 03 June 2021 MyLabX75VET GETTING STARTED 350080800 ...
Page 2: ......
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED RED Declaration of Conformity ...
Page 6: ...MyLab G E T T I N G S T A R T E D vi ...
Page 61: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 17 GETTING STARTED ...