S Y S T E M O V E R V I E W
MyLab
- G E T T I N G S T A R T E D
3 - 16
Error Messages
Whenever an internal fault occurs, the system automatically freezes and an
error message is displayed on the screen. Switch the system off and then turn
it on again to see whether the error message persists.
AO
Save anyway the log file (refer to the “Archive” section of the Advanced
Operation manual for further information) and contact the Esaote Service
department.
Errors in Battery Management
The battery icon is shown crossed out whenever an error in the battery
management occurs.
The number in the warning message indicates the type of error.
Error #1
This error indicates a fault on the Power Supply: in this case, information on
batteries may not be correct. The system displays the following message:
If this situation occurs, shut down
MyLab
, by keeping
O N /O FF
pressed, and
contact Esaote Service.
Error #2
This error indicates a failed access to the battery pack: in this case,
information on batteries may not be correct. The system displays the
following message:
If this situation occurs, shut down
MyLab
, by keeping
O N /O FF
pressed, and
contact Esaote Service.
Error #3
This error indicates that one battery couldn’t be charged. The system displays
the following message:
If this situation occurs, close the exam as soon as possible by pressing
EN D
EXAM
and switch the system off by pressing
O N /O FF
and then the main switch
Error #1: wrong communication with
power supply. The automatic shut down
is disabled.
Error #2: wrong communication with
battery logics. The automatic shut down
is disabled.
Error #3: problem with battery
charging
.
Summary of Contents for MyLabX6
Page 1: ...Rev 08 February 2020 MyLabX6 MyLabX7 GETTING STARTED 350031500...
Page 2: ...MyLab G E T T I N G S T A R T E D ii...
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED EC Declaration of Conformity...
Page 6: ...MyLab G E T T I N G S T A R T E D vi...
Page 7: ...MyLab G E T T I N G S T A R T E D vii GETTING STARTED RED Declaration of Conformity...
Page 8: ...MyLab G E T T I N G S T A R T E D viii...
Page 9: ...MyLab G E T T I N G S T A R T E D ix GETTING STARTED...
Page 10: ...MyLab G E T T I N G S T A R T E D x...
Page 16: ...I N D E X MyLab G E T T I N G S T A R T E D xvi...
Page 28: ...I N T R O D U C T I O N MyLab G E T T I N G S T A R T E D 1 12...
Page 49: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 13 GETTING STARTED...
Page 66: ...P R E P A R I N G T H E S Y S T E M MyLab G E T T I N G S T A R T E D 4 12...