Mode
PC m
ode
When the PC Mode key has been pressed, the screen
ap pears as shown at right. The ergometer is waiting for
commands from the external ECG unit.
With the arrrow keys, the saddle height can be electrically
adjusted on the ergoselect 200 (on the ergoselect 400,
these key adjust the height of the drive unit).
As soon as the ergometer receives commands from the
controlling ECG unit or PC, the exercise test will start and
the corresponding values will be displayed.
The exercise test can only be terminated with the corre‑
sponding command from the controlling ECG unit.
Note
• All functions are locked while the ergometer is operating
in PC mode, except for the saddle height adjustment and
the blood pressure key.
• To reactivate the saddle height adjustment function,
press
and the arrow keys will again be displayed.
• Additional blood pressure measurements can be initiated
with
.
Saddle
Saddle
PC Mode: Waiting for start
Exercise test running
d
isplay
during
exercise
test
1 most recent BP value (systolic/diastolic pressures) or
cuff pressure during inflation and bar graph indicating
microphone signal strength (see below)
2 SpO2 (%), heart rate (BPM)
3 duration of exercise test (minutes:seconds)
4 current load in watts
5 pedal speed (RPM)