5
Before use
The caregiver or user is responsible for making the following checks before use:
•
Personnel must have received qualified instructions in how to use the Luna.
•
The Luna is being used in track systems produced or approved by Winncare Nordic.
•
Installation of track system for the Luna is carried out by qualified installers approved
by dealer or Winncare Nordic.
•
Luna can only be fitted with spreader bars supplied by Winncare Nordic.
•
Luna has been developed for use with ErgoSlings and other sling products approved by
Winncare Nordic. See brochure for ErgoSlings.
•
The Luna is intended for use under normal temperatures (+2
˚C
/35.6
˚
F
to +40˚C
/104
˚
F)
and relative humidity (40 – 90%).
•
If the Luna is used near shower system or bath, make sure that the Luna or hand
control is not submerged (see section 3.3).
•
The Luna should never be used in an environment containing chlorine or in other
aggressive environments.
Safety checks
•
Check lifting strap for wear and tear or damage. If there is any sign of a defects, the
Luna must be taken out of use immediately.
•
Make sure the emergency stop has not been activated. If the red lamp is lit up, press
and hold the reset button for 5 seconds to release the emergency stop.
•
Check the up/down function, if any abnormal sound is heard the Luna must not be
used before a service check has been performed by a qualified technician.
•
Make sure that the strap can move freely into the Luna without twisting.
•
Check lifting sling for any damage especially with regards to lifting straps. The sling
must not be used if damage or defects are detected.
•
If in doubt on how to use the Luna, please contact the dealer or Winncare Nordic for
advice and guidance.
Terminology
Luna mounted in strap
Bayonet socket for spreader bar
Carry handle
Bayonet fixing point for Luna
Hook for sling lifting strap
Track system
Function buttons and handset connection
Strap
Summary of Contents for LunaIPx4