background image

80114-831

2019-05

QUICK GUIDE

IES

 

This quick guide is no substitute for briefing and user manual. This quick guide does not include any safety notes. Make yourself familiar
with the safety notes in the user manuals.

Necessary operating steps

Depending on which electrosurgical unit you are working with, certain
preparation steps are necessary.

Read the appropriate chapter in this User Manual: e.g. 

The IES 3 in

combination with a VIO 3 electrosurgical unit

.

Connect the desired accessories, e.g. prefilter, water trap, instru-
ment.

Compared to the 

OPEN mode

, the 

LAP mode

 has a lower suction capac-

ity with the same settings, and doesn’t have a turbo button. These fea-
tures of the 

LAP mode 

serve to protect the patient during laparoscopic

surgery and to support the stability of the pneumoperitoneum.

You can suction with the settings of the device or choose to change the
settings (see back).

Option 1

You use the IES 3 with an Erbe electrosurgical unit VIO 3, VIO D/S
series and the devices are connected to each other. 

You use the IES 3 with any desired electrosurgical unit and the au-
tomatic activation device is installed.

Activating the HF instrument.

The IES 3 starts automatically.

Option 2

You use the IES 3 without connection with an Erbe electrosurgical
unit and without an automatic activation device. 

Activate the IES 3 with the footswitch or with the activation but-
ton on the touchscreen of the device.

 

Activation screen: 

The segments on the right-hand side of the screen

show the course of the 

suction shut-off delay

. The segments on the

left-hand side of the screen show the course of the 

standby suction

time

 (not visible in the image).

Order of suction types

So long as you activate the HF instrument or press the footswitch or
activation button, 

suction

 is active. If you have selected times above

0 s in the 

Settings 

menu, the 

suction shut-off delay

 and the 

standby

suction time

 will follow respectively.

Observe the remaining filter display and the messages on the touch-
screen. 

!

1

Preparing the IES 3

2

Connecting the accessories

3

Switch on the device

4

Select the suction mode

5

Suctioning

6

Checking the filter capacity

Reviews: