4. 

Service and maintenance

4.1 

Maintenance / cleaning

The device needs no maintenance. Cleaning may be needed for the internal fans, the frequency of cleanse is 

depending on the ambient conditions. The fans serve to cool the components which are heated by the inherent 

power loss. Heavily dirt filled fans can lead to insufficient airflow and therefore the DC terminal would switch off 

too early due to overheating or possibly lead to defects.
Cleaning the internal fans can be performed with a vacuum cleaner or similar. For this the device needs to be opened.

4.2 

Fault finding / diagnosis / repair

If the equipment suddenly performs in an unexpected way, which indicates a fault, or it has an obvious defect, this 

can’t and must not be repaired by the user. Contact the supplier in case of suspicion and elicit the steps to be taken.
It will then usually be necessary to return the device to the supplier (with or without guarantee). If a return for 

checking or repair is to be carried out, ensure that:

• 

the supplier has been contacted and it is clarified how and where the equipment should be sent.

• 

the device is in fully assembled state and in suitable transport packaging, ideally the original packaging.

• 

a fault description in as much detail as possible is attached.

• 

if shipping destination is abroad, the necessary customs papers are attached.

4.2.1  Firmware updates

Firmware updates should only be installed when they can eliminate existing bugs in the firm

-

ware in the device or contain new features. 

The firmware of the control panel (HMI), of the communication unit (KE) and the digital controller (DR), if neces

-

sary, is updated via the rear side USB port. For this the software EPS Power Control is needed which is included 

with the device or available as download from our website together with the firmware update, or upon request.

However, be advised not to install updates promptly. Every update includes the risk of an inoperable device or 

system. We recommend to install updates only if...

• 

an imminent problem with your device can directly be solved, especially if we suggested to install an update 

during a support case

• 

a new feature has been added which you definitely want to use. In this case, the full responsibility is transferred 

to you.

Following also applies in connection with firmware updates:

• 

Simple changes in firmwares can have crucial effects on the application the devices are used in. We thus recom

-

mend to study the list of changes in the firmware history very thoroughly.

• 

Newly implemented features may require an updated documentation (user manual and/or programming guide, 

as well as LabView VIs), which is often delivered only later, sometimes significantly later

Summary of Contents for PSB 9000 3U Slave series

Page 1: ...E P S S t r o m v e r s o r g u n g Doc ID PSB9SEN Revision 01 Date 06 2018 Operating Guide PSB 9000 3U Slave Bidirectional DC Power Supply...

Page 2: ......

Page 3: ...AC supply 23 2 3 5 Connection to DC loads 25 2 3 6 Grounding of the DC terminal 26 2 3 7 Connection of remote sense 26 2 3 8 Connecting the Share bus 27 2 3 9 Connecting the USB port 27 2 3 10 Initial...

Page 4: ...ning and safety notices as well as general notices in this document are shown in a box with a symbol as follows Symbol for a life threatening danger Symbol for general safety notices instructions and...

Page 5: ...Slave Ancillary unit for master slave operation 3U 3 rack units of height Maximum current of the device in Ampere Maximum voltage of the device in Volt Series 9 Series 9000 Type identification PSB Po...

Page 6: ...urce ESD regulations must be applied when plugging interface cards or modules into the relative slot Interface cards or modules may only be attached or removed after the device is switched off It is n...

Page 7: ...e and cleaning of the equipment ensure that all employees who use the equipment have read and understood the manual Furthermore the users are to be regularly schooled in working with the equipment and...

Page 8: ...ng of the device DC terminal will be switched off Non critical Signal OVP OverVoltage Overvoltage shutdown of the DC terminal due to excess voltage entering the device or generated by the device itsel...

Page 9: ...urrent protection range 0 396 A 0 396 A 0 231 A 0 132 A Overpower protection range 0 16 5 kW 0 16 5 kW 0 16 5 kW 0 16 5 kW Temperature coefficient for set values K Voltage current 100 ppm Capacitance...

Page 10: ...3 of maximum current Insulation Allowed float potential shift on the DC terminal Negative terminal to PE Max 400 V DC 400 V DC 725 V DC 725 V DC Positive terminal to PE Max 400 V DC 400 V DC 1000 V D...

Page 11: ...on at 10 UAC 0 02 UMax 0 02 UMax 0 02 UMax 0 02 UMax Remote sensing compensation Max 5 UMax Max 5 UMax Max 5 UMax Max 5 UMax Voltage regulation power supply Load regulation at 0 100 IOUT 0 05 UMax 0 0...

Page 12: ...1800 V DC 1800 V DC 1800 V DC 1800 V DC AC input PE 2 5 kV DC AC input DC terminal 2 5 kV DC Miscellaneous Cooling Temperature controlled fans front side inlet rear exhaust Ambient temperature 0 50 C...

Page 13: ...1 570451 25 www eps germany de info eps germany de PSB 9000 3U Slave Series 1 8 4 Views Figure 1 Front view Figure 2 Rear view A Power switch D Share Bus and remote sensing connection G Matser slave p...

Page 14: ...tromversorgung GmbH Alter Postweg 101 86159Augsburg Germany Fon 49 821 570451 0 Fax 49 821 570451 25 www eps germany de info eps germany de PSB 9000 3U Slave Series Figure 3 Side view right Figure 4 S...

Page 15: ...Page 15 EPS Stromversorgung GmbH Alter Postweg 101 86159Augsburg Germany Fon 49 821 570451 0 Fax 49 821 570451 25 www eps germany de info eps germany de PSB 9000 3U Slave Series Figure 5 Top view...

Page 16: ...el For a detailed description see section 1 9 4 The control panel HMI 1 On Off button Can be used to switch the DC terminal on or off during manual operation while LED Remote off 2 USB port For quick...

Page 17: ...ote control when the unit is being used in stand alone operation The additional USB port on the front side is used for quick access to all the DC terminal related parameters and settings The configura...

Page 18: ...ff 1 9 4 2 USB port The front USB port is easier to access than the one on the rear side and intended for quick setup of DC terminal related values and settings Doing so is only necessary and possible...

Page 19: ...a two quadrants operation setup Following power supply and electronic load series are compatible on the so called Share bus PSB 9000 PSBE 9000 PSI 9000 2U 24U PSI 9000 3U Slave ELR 9000 ELR 9000 HP E...

Page 20: ...commissioning the equipment should be visually inspected for damage and completeness using the delivery note and or parts list see section 1 9 3 Scope of delivery An obviously damaged device e g loos...

Page 21: ...bination for models of PSB 9000 3U Slave series as it can be found in series PSB 9000 15U and PSB 9000 24U for example Advantage of this combination lower costs Disadvantages of this combination in ca...

Page 22: ...m current of the device Continuous operation at the approved limit generates heat which must be removed as well as voltage loss which depends on cable length and heating To compensate for these the ca...

Page 23: ...ollowing phases Nominal power Phases Supply type 15 kW or higher L1 L2 L3 PE Three phase The PE conductor is imperative and must always be wired 2 3 4 1 Cross section For the selection of a suitable c...

Page 24: ...the AC input con nection block on the rear It is used to prevent the AC plug from loosening and disconnecting due to vibrations or similar The fixture is also used as strain relief Being tied to the...

Page 25: ...parallel connection can be between 60 A two units and 5760 A 16 units From a certain current the handling of cables matching the current becomes impractical and the use of copper bars is required The...

Page 26: ...dius 2 3 6 Grounding of the DC terminal Grounding one of the DC terminal poles is allowed Doing so can result in a potential shift of the grounded pole against PE Because of insulation there is a max...

Page 27: ...lication it is recommended to disconnect the unit s from the Share bus because even when not powered they can have a negative impact on the control signal on the bus due to their impedance Disconnecti...

Page 28: ...n are suitable for your intended application of the device and adjust them if required as described in the manual In case of remote control via PC read the additional documentation for interfaces and...

Page 29: ...mode unloaded operation is not considered as a normal operation mode and can thus lead to false measurements for example when calibrating the device The optimal working point of the device is between...

Page 30: ...active the condition CC mode active will be indicated on the graphics display by the abbreviation CC and this message will be passed as a signal to the analog interface as well stored as status which...

Page 31: ...t depending on the input voltage according to Ohm s law IIN UIN RSET With series PSB 9000 3U master and PSB 9000 3U Slave the difference between external voltage supplied and internal set value determ...

Page 32: ...xternal voltage source to the DC terminal only the voltage set value determines the operation mode When connecting a load which cannot generate a voltage only source mode can be run Rules for applicat...

Page 33: ...device remains powered i e like after a temporary blackout can be set up via a specific command 3 4 2 Overtemperature An overtemperature alarm OT can occur from an excess temperature inside the devic...

Page 34: ...ice from providing an output voltage higher than 60 V However the alarm could also be trig gered by an external source providing an excess voltage to the DC input of the device A safety OVP alarm can...

Page 35: ...or operation LED Power changes to green There is a configurable option which determines the condition of the DC terminal after power up Factory setting here is OFF Changing it to Restore will cause th...

Page 36: ...ossible setting them only while the Slave device is not in control by a master device Outside of master slave the device could be controlled remotely with software EPS Power Control but also from cust...

Page 37: ...l RESistance MINimum SYSTem CONFig OPD SYSTem NOMinal VOLTage SYSTem CONFig OPD ACTion SYSTem SINK ALARm COUNt OCURrent SYSTem CONFig OPD ACTion SYSTem SINK ALARm COUNt OPOWer SYSTem CONFig OUTPut RES...

Page 38: ...n case the DC terminal is switched off and the LED Error is lit use button On Off 2 The LED should go off and with another push on On Off the DC terminal could be switched on again If the LED remains...

Page 39: ...egister can be read via USB With this series actions SIGNAL and WARNING are equal ALARM On reaching the condition which triggers the event with action ALARM a bit in the status register of the device...

Page 40: ...ues and status The Share bus is intended to balance the units dynamically in their power especially if the master unit runs a func tion In order for this bus to work correctly at least the DC minus po...

Page 41: ...custom software The programming guide as included in the delivery on USB stick explains remote configuration for master slave in custom applications 3 8 1 6 Operating the master slave system After the...

Page 42: ...rm they report the alarm to the master In situations where the current is probably not balanced between the units it can occur that one unit generates an OC alarm though the global OC limit of the MS...

Page 43: ...transport packaging ideally the original packaging a fault description in as much detail as possible is attached if shipping destination is abroad the necessary customs papers are attached 4 2 1 Firmw...

Page 44: ...acturer No RMA number is needed It is sufficient to package the equipment adequately and send it together with a detailed description of the fault and if still under guarantee a copy of the invoice to...

Page 45: ......

Page 46: ......

Reviews: