Cleaning Instruction:
All blood and other body fluids must be thoroughly cleaned from surfaces and
objects before disinfection by the germicidal wipe.
Open, unfold and use first germicidal wipe to remove heavy
soil.
Disinfection Instruction
:
Unfold a clean wipe and thoroughly wet all the surface of the meter, including the
strip port and other connection port. Unfold a clean wipe and thoroughly wet all the surface of the lancing
device, including cap or AST cap if used. Treated area must remain visibly wet for a full 2 minutes. Let the
devices air dry for 0.5 minute. Do disinfection once per week.
Do Not:
Get any moisture in the test strip slot.
Spray any cleaning solution directly onto the
meter.
Put the meter under water or liquid.
Pour liquid into the meter.
Do:
Make sure the meter is turned off during cleaning
and disinfection.
Keep the test strip vials tightly closed when
performing the cleaning and disinfection
procedures because the fumes from the
disinfectant may affect the performance of the strip
After cleaning or disinfection, please perform the
physical appearance and performance check of
devices.
Please follow the instruction on page 44 to 46.