46
2-2. Performance check of the Lancing device
Step
Check item
Accept
Result
After each cleaning
1. Cap the normal cap to the lancing device.
Can the cap be screwed on the lancing device?
Yes
2. Pull on the sliding barrel and then release. Can the lancing device click?
Yes
3. Press trigger button.
Can the lancet be fired?
Yes
After each disinfection
1. Install the lancet on the lancet holder.
Can the lancet be fixed on the lancet holder?
Yes
2. Cap the AST cap to the lancing device.
Can the AST cap be screwed on the lancing
device?
Yes
3. Remove the AST cap. Recap the normal
cap to the lancing device.
Can the cap be screwed on the lancing device?
Yes
Can the penetration depth ring be adjusted?
Yes
4. Pull on the sliding barrel and then release. Can the lancing device click?
Yes
5. Press trigger button.
Can the lancet be fired?
Yes
Action:
If any of the results is “No”, please call the Customer Care Service toll-free at 866-994-3345(Eastern
Time, Mon-Fri 9:00AM~5:00PM) for replacement with a new device.