Pulse oximeter MySign
®
S
8.1 Replacing the battery
Only batteries (Part No. 1001734) of the following type may be used:
•
Li-ion 3.6 V DC
•
2900 mAh
To replace a faulty battery, remove the screw
c
and push down the cover
d
. Now take off the cover and remove
the battery connection cable
e
and the battery
f
.
The MySign
®
S may not be operated with regular, non-rechargeable batteries.
The battery connection cable has reverse polarity protection. The unit settings (such as date/time)
are saved for at least 5 minutes while the battery is being replaced. After this time, all settings
must be reconfigured.
Disposing of the unit, SpO2 sensor, battery
Do not dispose of the unit, SpO2 sensor or battery as household waste. Please return all of these to EnviteC
clearly labelled “Entsorgung (Please dispose of)”.
EnviteC-Wismar GmbH
Alter Holzhafen 18
23966 Wismar, Germany
Germany
Danger of explosion!
Do not throw batteries into fire or force them open.
8.2 Replacing the sensor
1. Remove the sensor
2. Disposing of the sensor
3. Connect a new sensor and check its functions
Note the arrow on the sensor plug and the unit socket to ensure the correct connection orientation.
Otherwise the contacts could be damaged and correct functioning cannot be guaranteed!