6.7.5 Application indicator
Indicates how long the SpO2 sensor has been used at the same application point (e.g. index finger).
The application indicator is given in hours from 0 - 8. The display starts as soon as the sensor is placed or re-
applied. This function can be switched on or off under “Application indicator”.
Select “Application indicator” from the “Alarm settings”
menu.
This is switched on or off with the right function key.
The symbol flashes if more than 4 hours have passed. After 8 hours, the following symbol
is shown.
The time is only reset to zero by the error message “Sensor off”.
Check sensor and application at least every 4 hours for proper functioning (check position and
patient skin for damage), and reposition if necessary. For older people, it is recommended that
checks are performed after just 2 hours to prevent bruises from forming.
Switching the unit off and back on interrupts the time measurement, which begins again at 0 hours.