12.3 Cleaning of tubes
•
Remove the tube from the device.
•
Rub the tubes with a moist cloth (soapsuds). NEVER dip the tube
into any fluid.
12.4 Maintenance and validity check prior to any use
•
Prior to using the device, manually check the device for any
mechanical damage.
•
If you notice any damage or functional errors, which may endanger
the safety of the patient and user of the device, then you should
only use the device again after it has been repaired.
Measurement technical check
The PhysioQuant is a measuring device in accordance with the German
'Medical Device Code - Operation V' ('MP Betrieb V') § 11/ Appendix 2.
As a result, the device must undergo a 'measurement technical check'
every 2 years. The first of these checks should take place on the date
indicated on the calibration mark.