9 Error
codes
E 03
Internal hardware error (contact the Service Department).
E 04
Batteries are empty. This is displayed as soon as the batteries have
insufficient capacity to carry out measurements.
E 05
Measurement time has elapsed. This is displayed when measurement
time (without inflation time) exceeds 60 seconds.
E 06
This error code will be displayed when:
•
the actual cuff pressure exceeds the maximum allowable inflation
pressure of 280 mmHg;
•
the device is not inflating beyond the set maximum pressure and is
waiting for the next measurement time.
E 07
The maximum inflation time of 60 seconds has elapsed. Error refers to
a loose cuff or tube, or another faultysealing.
E 08
Insufficient oscillations have been detected:
At least 8 oscillations need to be detected for a correct measurement.
The cuff should be attached sufficiently tightly (you should be able to
put one finger beneath the cuff, but not two). PhysioQuant will regulate
the deflation rate for follow-up measurements accordingly.
E 10
Memory full. 200 blood pressure measurements have been carried out
and the memory of the device is full.
E 11
Movement during diastole recognition.
E 12
Diastole is outside of the measurement range.
E 20
Systole is outside of the measurement range.
(E12 and E20 are displayed when the calculated systole or diastole
values are outside the range in which the oscillations are recorded.)
E 21
Systole is below the measurement range.
E 22
Systole is above the measurement range.
E 24
Difference between systole and diastole is too small (10 mmHg or less)
E 23
Movement during systole recognition.
Air deflation speed is too high, e.g. as a result of leakage.