Taking Sample(s) Using the 

Single Sample 

Mode 

1.

 

Place the pump in the desired location.  

2.

 

Scroll to 

Single Sample

 and press the 

F2 

button.    

3.

 

Place the sampling cassette onto the cassette retainer so that it fits snugly. Then either press the 

F1

 

button or wait a few minutes until the next screen appears.  

4.

 

Set the 

Run Time

; sampling run time can be set from 1 minute up to 9 hours, 59 minutes. 

5.

 

Press 

F1 

button. 

Run Time

 will count up to the chosen 

Set Time

; green LED will flash when sampling. 

Once complete, 

Sample Complete

 will be displayed, green LED will be constant, and unit will beep three 

times.  

6.

 

Press 

F2

 button to return to the main user screen. Remove cassette from pump by gently pulling 

upward. Seal cassette with tabs, document the sample run time, and send to laboratory for analysis.   

 

Taking Sample(s) Using the 

Sequential Sampling 

Mode 

1.

 

Place the pump in the desired location.  

2.

 

Scroll to 

Sequential Sample

 and press 

F2

 button.   

3.

 

Place the sampling cassette onto the cassette retainer so that it fits snugly. Then either press the 

F1

 

button or wait a few minutes until the next screen appears.  

4.

 

Enter the 

#On/Off Cycles

 (one cycle

 

run time followed by one cycle off time). Enter the 

Cycle Run Time

 

(run time will be the same for all run cycles). Enter the 

Cycle Off Time

 (off time will be the same for all 

off cycles). 

5.

 

Press 

F1 

button. 

Run Time

 of 

Cycle 01

 will count down; green LED will flash when sampling. Once 

complete, pump will switch to the 

Cycle 02

 

Off Time

 and count down; green LED will continue to flash 

during the off times. During the 

Off Time

, the cassette should be removed and a new cassette loaded 

onto the cassette retainer. Seal the sampled cassette with tabs, and document the sample run time. 

After each 

Off Time

, the RFID reader will detect tag on cassette if present and give one of the messages 

(see #5 above). 

Sample Complete

 message will only show after the entire 

Run Cycle

 is done (after the 

last cycle run time).  

6.

 

Press 

F2 

button

 

to return to the main user screen.  

    

Section 9 -

 

Retrieval of Event Records with App 

 

1.

 

Press the 

Bluetooth

 button on pump while in the main menu screen, 

Bluetooth icon

 will appear  

next to battery level indicator and icon will be gray.  

2.

 

Pair pump to your mobile device. Once paired, the icon will turn from gray color to blue color.   

3.

 

Initiate data transfer from mobile app. User screen will display picture of IAQ 15 Connect and a mobile 

device with arrows between during the data transfer. Message displayed, 

Keep your device connected 

with Bluetooth until all data is transferred

.  

 

Summary of Contents for IAQ 15 Connect

Page 1: ...T 1 800 343 5319 1 843 881 6560 W environmentalexpress com LA03115 Rev 1 IAQ 15 Connect Operating Instruction Manual...

Page 2: ...ncluded 5 Section 5 Physical Description 6 Section 6 Prior to Operation 7 Section 7 Calibration 7 Section 8 Sampling Modes 8 Section 9 Retrieval of Event Records with App 10 Section 10 Battery Status...

Page 3: ...ble for any consequential special indirect or compensatory damages arising from use of or in conjunction with this product The maximum liability of Environmental Express whether by reason of breach of...

Page 4: ...S s Operation is subject to the following two conditions 1 This device may not cause interference 2 This device must accept any interference including interference that may cause undesired operation o...

Page 5: ...in a manner for which it is not intended Technical Specifications Flow rate 15 LPM Accuracy 5 Housing PC ABS plastic Battery rechargeable nickel metal hydride NiMH 9 6 V 2500 mAh Run time 99 five min...

Page 6: ...ner 2 LCD 3 F1 button increase 4 F2 button enter 5 F3 button decrease 6 Bluetooth button 7 Power On Off and Wake up button 8 AC charger input jack Bottom View of Pump 1 Battery pack Battery cover 2 Gr...

Page 7: ...very day of use A flow indicator for Air O Cell cassettes is supplied with the pump for ease of calibration You may also calibrate the pump using an optional TSI 4046 Primary Calibrator with attachmen...

Page 8: ...tures a choice of 3 different sampling modes We strongly recommend that the sampling mode is chosen after calibration is complete Quick Sample This mode allows easy sampling with the most commonly use...

Page 9: ...llowed by one cycle off time Enter the Cycle Run Time run time will be the same for all run cycles Enter the Cycle Off Time off time will be the same for all off cycles 5 Press F1 button Run Time of C...

Page 10: ...10 environmentalexpress com 4 Once complete check mark will appear between the IAQ 15 Connect and mobile device with message stating Bluetooth Data Transfer Success...

Page 11: ...be displayed on the home screen Previous sample may be continued or a new sample may be started To charge the battery pack plug in the AC charger on the right side of the pump The yellow LED will fla...

Page 12: ...Section 13 Maintenance Proper care and maintenance of your IAQ 15 Connect pump is essential for a long operational life It is a delicate electronic device and should be treated accordingly Rough and a...

Page 13: ...art Heavy duty tripod sampling stand ZA0043 Remote extension tubing 5 ft ZBP 310 Calibration kit with TSI 4046 Primary Calibrator ZBP 200 CAL Flow indicator for calibrating Via Cell cassettes ZBP 305...

Reviews: