8 |
environmentalexpress.com
4.
Press
F2
button.
5.
The pump will begin running at a default flow rate; the green LED will begin flashing. Use the
F1
button
to adjust the flow down or the
F3
button to adjust the flow up until “15 LPM” registers on the flow
indicator or on the TSI calibrator’s LCD.
6.
Press the
F2
button to set the calibration and then
Calibration Complete
will appear; the green LED will
be constant. Press
F2
button to return to main menu screen.
Note:
When you enter the
Calibration
mode, the pump will reset its flow rate to the default setting. To check
the pump calibration without resetting the flow rate, you may use the
Verify Calibration
function.
Verifying the Calibration
1.
Press the
On/Off
button to turn the pump on.
2.
Scroll and select the
Verify Calibration
mode.
3.
Connect the flow indicator or TSI 4046 Primary Calibrator.
4.
Press
F2
button.
5.
Measure the flow rate; the green LED will begin flashing. Press
F2
button to stop.
Verification Complete
will be displayed on the screen and green LED will be constant. The calibration flow cannot be adjusted
during the
Verify Calibration
mode, it can only be checked.
Section 8 -
Sampling Modes
The IAQ 15 Connect pump features a choice of 3 different sampling modes. We strongly recommend that the
sampling mode is chosen after calibration is complete.
•
Quick Sample
— This mode allows easy sampling with the most commonly used sampling times of
1, 2, 5, or 10 minutes.
•
Single Sampling
— This mode is designed to take one sample at a predetermined amount of time.
The sampling time can be set anywhere from 1 minute to 9 hours, 59 minutes.
•
Sequential Sampling
— This mode allows the pump to be programmed to turn on and off at
predetermined times. The user programs the number of on/off cycles, the cycle run time, and the cycle
off time.
Taking Sample(s) Using the
Quick Sample
Mode
1.
Place the pump in the desired location.
2.
Scroll to
Quick Sample
and press the
F2
button.
3.
Place the sampling cassette onto the cassette retainer so that it fits snugly. Then either press the
F1
button or wait a few minutes until the next screen appears.
4.
Scroll and select the desired sampling time from 1, 2, 5, or 10 minutes.
5.
Press
F1
button.
Run Time
will count up to the chosen sampling
Set Time
; green LED will flash when
sampling. Once complete,
Sample Complete
will be displayed, green LED will be constant, and unit will
beep three times.
6.
Press
F2
button to return to the main user screen. Remove cassette from pump by gently pulling
upward. Seal cassette with tabs, document the sample run time, and send to laboratory for analysis.