The IAQ 15 Connect is an advanced portable, battery-powered air sampling pump designed for  

exclusive use with Air-O-Cell

®

 and Via-Cell

®

 cassettes for fast sample collection. Protection provided 

within the system may be impaired if the device is used in a manner for which it is not intended. 

 

Technical Specifications 

 

Flow rate: 15 LPM 

 

Accuracy: ±5% 

 

Housing: PC/ABS plastic 

 

Battery: rechargeable nickel metal hydride (NiMH), 9.6 V, 2500 mAh 

 

Run time: 99 five-minute samples on new, fully charged battery 

 

Dimensions:  3.75” W x 2.55” H x 6.85” D (9.5 x 6.5 x 17.4 cm) 

 

Weight: 1.25 lb (0.57 kg) 

 

Temperature: 41 to 104

°

F (5 to 40

°

C) 

 

Max humidity: 80% RH up to 88

°

F (31

°

C), decreasing linearly to 50% RH at 104

°

F (40

°

C) 

 

Altitude: up to 6561 ft (2000 m) 

 

Ingress protection: IPX0 

 

Use: indoor and outdoor 

 

Pollution degree: 2 

 

Electrical rating: 15V 

, 1.2A   

 

Section 4 -

 

What Is Included 

 

 

 

 

IAQ 15 Connect pump (Part # 

W30162

 

Air-O-Cell cassettes, 10/pk 

 

Air-O-Cell flow indicator 

 

Battery pack 

 

Battery cover key (Part # W40132) 

 

AC power adapter 

 

Carrying case 

Summary of Contents for IAQ 15 Connect

Page 1: ...T 1 800 343 5319 1 843 881 6560 W environmentalexpress com LA03115 Rev 1 IAQ 15 Connect Operating Instruction Manual...

Page 2: ...ncluded 5 Section 5 Physical Description 6 Section 6 Prior to Operation 7 Section 7 Calibration 7 Section 8 Sampling Modes 8 Section 9 Retrieval of Event Records with App 10 Section 10 Battery Status...

Page 3: ...ble for any consequential special indirect or compensatory damages arising from use of or in conjunction with this product The maximum liability of Environmental Express whether by reason of breach of...

Page 4: ...S s Operation is subject to the following two conditions 1 This device may not cause interference 2 This device must accept any interference including interference that may cause undesired operation o...

Page 5: ...in a manner for which it is not intended Technical Specifications Flow rate 15 LPM Accuracy 5 Housing PC ABS plastic Battery rechargeable nickel metal hydride NiMH 9 6 V 2500 mAh Run time 99 five min...

Page 6: ...ner 2 LCD 3 F1 button increase 4 F2 button enter 5 F3 button decrease 6 Bluetooth button 7 Power On Off and Wake up button 8 AC charger input jack Bottom View of Pump 1 Battery pack Battery cover 2 Gr...

Page 7: ...very day of use A flow indicator for Air O Cell cassettes is supplied with the pump for ease of calibration You may also calibrate the pump using an optional TSI 4046 Primary Calibrator with attachmen...

Page 8: ...tures a choice of 3 different sampling modes We strongly recommend that the sampling mode is chosen after calibration is complete Quick Sample This mode allows easy sampling with the most commonly use...

Page 9: ...llowed by one cycle off time Enter the Cycle Run Time run time will be the same for all run cycles Enter the Cycle Off Time off time will be the same for all off cycles 5 Press F1 button Run Time of C...

Page 10: ...10 environmentalexpress com 4 Once complete check mark will appear between the IAQ 15 Connect and mobile device with message stating Bluetooth Data Transfer Success...

Page 11: ...be displayed on the home screen Previous sample may be continued or a new sample may be started To charge the battery pack plug in the AC charger on the right side of the pump The yellow LED will fla...

Page 12: ...Section 13 Maintenance Proper care and maintenance of your IAQ 15 Connect pump is essential for a long operational life It is a delicate electronic device and should be treated accordingly Rough and a...

Page 13: ...art Heavy duty tripod sampling stand ZA0043 Remote extension tubing 5 ft ZBP 310 Calibration kit with TSI 4046 Primary Calibrator ZBP 200 CAL Flow indicator for calibrating Via Cell cassettes ZBP 305...

Reviews: