CALIBRATION PROCEDURES
H
2
S Gas Calibration
Depending on the application, the 330S-Ex and 330SDS-Ex H
2
S analyzer will
require more or fewer periodic calibrations. There is no specific time as to
how often the H
2
S analyzer should be calibrated. It will depend on the
application, importance of accuracy for the application, and how dirty or clean
the environment and sample are. The following is the calibration procedure
for the 330S-Ex and 330SDS-Ex H
2
S analyzer:
1. Source a calibration gas of H
2
S in balance of N2 regulated to 15 psi
(check expiry date). H
2
S concentration to be approximately 2/3 of full-
scale range or close to the H
2
S alarm set point.
2. Press the bypass button and verify the "Bypass" LED illuminates (alarms
will be held in the non- alarm state).
3. Turn off all gas supplies to the analyzer and check that a sufficient
amount of H
2
S sensing tape is installed.
4. Press the "Menu/Set" button until "Mtr Run" is displayed. Press the right
arrow [
→], the
H
2
S sensing tape will advance for approximately 10
seconds.
5. Press the "Menu/Set" button until "mV" is displayed ("###mV"). If the mV
reading is 1000mV (+100mV), proceed to the next step, otherwise re-
zero sensor (Refer to "Re-zero Sensor Procedure" on page 33).
6. Connect calibration gas to calibration port and turn the 3-way calibration
valve 180°. The valve handle should be pointing towards where the gas
bottle tubing is connected to (Calibration Inlet).
7. Turn on sample inlet valve, ensure that the sample regulator is supplying
15psig to the eductor (make sure there is suction from the eductor
block). Adjust the flow meter to 2.0. Wait until the H
2
S reading has
stabilized (10 to 15 minutes).
8. With calibration gas applied, if H
2
S reading is satisfactory (+2% of
analyzer full range) skip to step 16, if H
2
S reading is not satisfactory a
gain adjustment is required, continue to step 9.
9. Press the "Menu/Set" button until the gain setting is displayed ("###
Gain").
10. Calculate the new gain. New gain value should be within approximately
25% of the gain installed at the factory.
Summary of Contents for 330S-Ex
Page 2: ......
Page 55: ...Envent Engineering Ltd Page 50 Revision 1...
Page 56: ...Envent Engineering Ltd Page 51 Revision 1...
Page 57: ...Envent Engineering Ltd Page 52 Revision 1...
Page 58: ...Envent Engineering Ltd Page 53 Revision 1...
Page 59: ...Envent Engineering Ltd Page 54 Revision 1...
Page 60: ...Envent Engineering Ltd Page 55 Revision 1...
Page 61: ...Envent Engineering Ltd Page 56 Revision 1...
Page 62: ...Envent Engineering Ltd Page 57 Revision 1...
Page 63: ...Envent Engineering Ltd Page 58 Revision 1...
Page 64: ...Envent Engineering Ltd Page 59 Revision 1...
Page 65: ...Envent Engineering Ltd Page 60 Revision 1...
Page 66: ...Envent Engineering Ltd Page 61 Revision 1 Safety Data Sheet for H2S Sensing Tape...
Page 67: ...Envent Engineering Ltd Page 62 Revision 1...
Page 68: ...Envent Engineering Ltd Page 63 Revision 1...
Page 69: ...Envent Engineering Ltd Page 64 Revision 1...