STACK 602
9 Maintenance
ENVEA
UK Ltd
(PC-000848-MA Issue 5, December 2021)
vii
5.2.4
Setting up the Air Purges
Air purges are normally fitted to the Transmitter and Receiver to minimise
contamination of the lenses. These are supplied as a filter/regulator assembly
correctly fitted to the Transmitter and Receiver units.
The air purges require a supply of dry, clean (oil-free) instrument-quality air, which
should be set to deliver a flow rate of 30–60 l/min. Connection to the air purge is
by a ¼” BSP fitting.
NOTE
: It is recommended to adjust the air flow before
fitting the sensors to the stack or
ductwork.
5.3
Connecting the Transmitter (Tx)
Figure 12: Transmitter (Tx) PCB connections.