8 of 32
EN109-1730752-41
CHAPTER 4 - SAFETY
4.1 General introduction
It is important to read the instructions carefully before using Tecarpuls. Please make sure the
instructions are available to all users.
1.
The Tecarpuls should only be connected to an approved electrical installation, with a grounded
plug.
2.
Make sure the electrode cables are firmly seated in the connectors on the device itself. Screw
the mounting ring of the connector of the treatment probe cable into the connector on the
right side of the device. Plug ground electrode (metal rod) fully into the connector on the right
side of the device.
3.
Make sure the ground electrode cable is fully seated inside the ground electrode itself (metal
rod).
4.
If you need to remove the cable from the ground electrode, or from the treatment probe: First
turn off the power to the device and then unplug the plug from the wall.
5.
Never pull directly the cables themselves to pull them out of the electrodes or device. Then
you can damage the electrical inner wire and the device might not work. Always pull out wires
by grasping the contacts themselves and pulling gently.
6.
The device must not be used in the vicinity (less than 2 m) of other electrotherapy devices in
use. By using the device near (< 2 meters) of short wave or microwave equipment, this can
result in instability in the performance of Tecarpuls.
7.
Other electrotherapy / electrical devices should NOT be used simultaneously on the patient
when treated with Tecarpuls. It can cause electrical interference and result in electrode burns.
8.
To prevent electromagnetic interference, we strongly recommend the use of separate main
power connection lines for Tecarpuls and other therapeutic devices
9.
The device must not be used in so-called hydrotherapy rooms (wet rooms)
10.
Wireless phones should not be used near the machine.
11.
Always apply a little cream to the earth electrode, but be more generous in the treatment
area.
Use only
original Electrode Cream
, as this has been developed for this device. Other type of
cream may cause lack of effect / cause harm to the patient.
12.
Communicate with the patient during treatment, and encourage the patient to notify if the
heat becomes uncomfortable. Then the strength is reduced 1-2 steps.
13.
Avoid touching the patient during treatment.
The patient can then feel for a moment a slightly stinging local feeling.
Summary of Contents for Tecarpuls
Page 34: ...EN109 1730752 41 INSTRUCTIONS FOR USE 31 March 2020 1730752 41 2460 TECARPULS ...
Page 66: ...DE109 1730752 41 GEBRAUCHSANWEISUNG 31 März 2020 DE109 1730752 41 2460 TECARPULS ...
Page 98: ...FR109 1730752 41 MANUEL D UTILISATION 31 Mars 2020 FR109 1730752 41 2460 TECARPULS ...
Page 162: ...IT109 1730752 41 ISTRUZIONI PER L USO 31 Marzo 2020 IT109 1730752 41 2460 TECARPULS ...