10.1.3 Vacuum electrodes
There is a choice of large and small electrodes. The areas of the electrodes correspond to those
of the 4 x 6cm and 6 x 8cm flexible rubber electrodes. The vacuum electrodes are sufficiently
flexible to ensure optimum contact with the skin, but rigid enough to prevent any changes in the
contour of the part being treated, allowing full advantage to be taken of the massage effect of
the pulsed vacuum.
Keep the sponge pads well moistened during treatment.
After use clean the sponge pads as described in the User Maintenance instructions.
10.1.4 Self-adhesive electrodes
Self-adhesive electrodes have higher series impedance then flexible rubber electrodes. This
can cause the stimulator to terminate treatment at higher current amplitudes. When this occurs
it is recommended to continue the treatment with flexible rubber electrodes, combined with
properly moistened sponge pads.
Self-adhesive electrodes are not recommended for use with currents that contain a DC
component.
CAUTION: Do not use electrodes on open wounds.
10.1.5 Electrolytic effects
Electrolysis occurs under the electrodes when current types with a DC component are applied.
Because the greatest concentration of electrolytic by-products caused by ion migration occur
under the electrodes, we recommend the use of the supplied sponges to keep the effects to a
minimum. Make sure that the sponges are kept well moistened and place the thick side of the
sponge between the flexible rubber electrodes and the patient.
10.1.6 Current density
In the particular standard for Electrical Nerve and Muscle Stimulators, IEC 60601-2-10, it is
recommended not to exceed a current density of 2 mA r.m.s. / cm², otherwise skin irritations or
burns can occur. For current types that contain a DC component we recommend not to exceed
a current density of 0.2 mA / cm².
To find the maximum recommended current amplitude in mA for the Interferential,
Premodulated and Russian Stimulation current waveforms, multiply the electrode surface in cm²
by two. For all other current waveforms the stimulator output current can never exceed 50 mA
r.m.s. This implies that with an electrode surface of 25 cm² the current density can never exceed
2 mA r.m.s. / cm². As a rule of thumb for smaller electrodes, such as the 3.2mm self adhesives,
the maximum current setting available on the stimulator for a given current waveform should
proportionally be reduced.
For a precise calculation of the r.m.s. value of a pulsed current waveform the following formula can be used:
I
RMS
= I
peak
√ ( Phase duration [µs] * pulse frequency [Hz] * 10
6
)
For symmetrical TENS currents, the Phase duration should be multiplied by 2. The value of the peak current I
peak
can be taken from the current display.
Electrodes should be placed with care, ensuring good electrical contact over the entire electrode
surface.
Summary of Contents for 1600905
Page 1: ...6 series User Manual ...
Page 3: ...6 series User Manual ...
Page 4: ......
Page 5: ...3 1600751 4D ...
Page 90: ...88 1600751 4D ...
Page 92: ...90 1600751 4D 13 6 3 LP 13 6 4 CP 13 6 5 CPid ...
Page 95: ...93 1600751 4D 13 7 7 IG150 Monophasic Envelope 13 7 8 IG150 Biphasic Envelope ...