3.0 Installation
3.1 Mounting of Instrument
The
M
ED
A
IR
2200
should be located near the pipe or tank containing the air to be monitored, and upstream from where the air is
being used. The
M
ED
A
IR
2200
must be installed such that it samples the compressed air before it reaches the users.
Mount the instrument on an appropriate vertical surface using the mounting flanges provided. Avoid areas with excessive vibration
or temperature extremes. The holes in the flanges are 0.31 inch in diameter and form a 6 x 10.75 inch rectangle.
See Figure 3.
It is recommended to use #8 drywall anchors and screws for mounting the
M
ED
A
IR
2200
to a drywall/sheetrock surface.
Dimensions are in inches.
Figure 3: M
ED
A
IR
2200
Mounting Dimensions
Mounting Holes
0.31” dia. 4 places