
MedAir
ENMET
Manual Release Date – June 9, 2017
P a g e
| 5
Manual Part No. – 80002-042
Figure 1: External Features of the M
ED
A
IR
2200
* N
OTE
:
Typical gas reading & alarm locations, depending on instrument configuration, alarms & readings may be in alternate
locations
Sample Air Hose, 5ft Supplied with Monitor
Example of Regulator, Not Supplied with Monitor
Figure 1A: Sample Air Hose and Regulator
N
OTE
:
Typical facility airline pressures range from about 55 to 125
PSI
. The outlet pressure of the
regulator can be connected to the inlet of the
M
ED
A
IR
2200
and should be set at 55
PSI
,
±
5
PSI
.
Regulator not supplied with the monitor
O
2
%
CO
ppm
Front Cover
Latch
Male Quick Release Fitting
Input Port
Sample/Calibration
See note below
Audio Alarm
Fault LED
SELECT
Switch
Power LED
OPTION
Switch
A
LARM
A
CKNOWLEDGE
/
A
LARM
D
EFEAT
Switch
*Visual Alarm O2
(If Oxygen option is installed)
See *note below
*Visual Alarm CO
(See *note below)
Mounting Flanges
2 places
Dew
Point
*Visual Alarm
Dew Point
(If Dew Point option is
installed)
See *note below
Display, see *note below
CO reading
O
2
reading
Dew Point reading
Channel #1
Channel #3
Channel #4
Channel #2
Visual Alarm 2
ALARM
Female Quick Release
to Inlet Port to M
ED
A
IR
2200
¼”
N
OTE
:
When connecting to a standard 55
PSI
USA Medical air system, Regulator is Not required.
Male¼” NPT Fitting
To Air System
Female Quick Release
to Sample Port