ENMET MED AIR 2000 Operation And Maintenance Manual Download Page 10

ENMET

 Corporation

M

ED 

A

IR 

2000

6

3.2 Sample Air Supply

Tap the pipe or tank containing the breathing air and use appropriate fittings to connect the sample input hose.  The
instrument is designed to operate from an air supply pressure of 50-55 

PSIG

; adjust the regulator and set the pressure at

55 

PSIG

.  The flowmeter on the display panel indicates approximately 2 SCFH when the input pressure is 55 

PSIG

.

The sample air exits the instrument from two separated ports on the back surface of the enclosure.  Take care not to
obstruct these exit ports.  After mounting the enclosure, they are not accessible.
Be sure that the red sample-calibrate valve handle on the right side of the enclosure is pointed down toward the sample
input port.

3.3 Power Supply

Plug the line cord in a source of 110

V

AC

 power.  The input power can vary from 100 to 240

V

AC

, 50/60 Hz; if other

than 110 

V

AC

 power is desired, the plug on the line cord must be changed.

Upon supplying air and power to the instrument:

ƒ

 

The green power on LED is lit.

ƒ

 

The display backlight is lit, and numbers are given on the display.

The instrument may go into alarm briefly, but the sensors stabilize quickly.  If the instrument persists in alarm,
acknowledge the alarm by pressing the 

ALARM ACKN/AUDIO DEFEAT

 switch.  If alarm persists longer than 30

minutes, call 

ENMET

 customer service personnel.

3.4 Outputs

Two types of alarm outputs are available, relay contacts and optional 4-20mA outputs.

CO

O2

DP

N/A

Fault

LOOP

Figure 2A: Terminal Strip

3.4.1 Relay Contacts

Relay contacts are available for each alarm; these are 

SPDT

, rated at 2.0Amp at 110

V

AC

, and may be latching or non-

latching as required by the application.  They are accessed on the terminal strip at the bottom of the circuit board see

Figure 2 & 2A

.  The positions given in Table 1:

Table 1: Relay Contacts

 

Position

 

Function

 

Contact

 

1

 

CO Alarm

 

NC

 

2

 

CO Alarm

 

C

 

3

 

CO Alarm

 

NO

 

4

 

O

2

 Alarm*

 

NC

 

5

 

O

2

 Alarm*

 

C

 

6

 

O

2

 Alarm*

 

NO

 

7

 

DP Alarm

 

NC

 

8

 

DP Alarm

 

C

 

9

 

DP Alarm

 

NO

 

10

 

Not Used

 

 

11

 

Not Used

 

 

12

 

Not Used

 

 

* This relay is activated by both the deficiency and abundance alarms.

These relay coils are energized when they are in the non-alarm state; the contact conditions given above are for the
non-energized state, which is identical to the alarm state.

1

2

 3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18 19 20  21  22  23

 

NC

  

C

 

NO NC C

NO NC

 

C NO

 

NC

  

C

 

NO

 

NC

  

C

NO

 

G

 +

G

 +  

G

  +   

G

Summary of Contents for MED AIR 2000

Page 1: ...MCN 167 09 20 00 MCN 176 03 06 98 MCN 200 09 04 98 MCN 204 01 29 99 MCN 237 09 21 00 MCN 274 05 01 02 MCN 327 02 03 05 MCN 332 04 15 05 MED AIR 2000 Operation and Maintenance Manual Andersen Medical...

Page 2: ......

Page 3: ...peration Condition 8 4 2 Alarm Set Points 8 4 3 Alarm Latching 8 4 4 Audio Defeat 9 4 5 Display 9 4 6 Operational Menu 9 4 7 Fault Indications 10 4 7 1 Low Flow Indication 10 4 7 2 Other Fault Indicat...

Page 4: ...SPLAY 8 TABLE 4 FACTORY ALARM SET POINTS 8 FIGURE 5 MED AIR 2000 OPERATION MENU FLOW CHART 9 FIGURE 6 MED AIR 2000 MAINTENANCE MENU FLOW CHART 11 FIGURE 7 CONNECTION OF CALIBRATION GAS CYLINDER 13 TAB...

Page 5: ...ected verified and properly packaged at our company and delivered to the carrier in good condition When it was picked up by the carrier at ENMET it legally became your company s property If your shipm...

Page 6: ...ord A cord to supply 110 VAC to the equipment Not illustrated Audio Alarm A loud horn activated by certain alarm conditions Mounting Flanges Flanges with holes for mounting the enclosure to a vertical...

Page 7: ...put See note below Audio Alarm Flow Meter Strain Relief for Line Cord Fault LED SELECT Switch Power LED OPTION Switch ALARM ACKNOWLEDGE ALARM DEFEAT Switch Visual Alarm O2 If Oxygen option is installe...

Page 8: ...ld Housing The sample manifold the carbon monoxide and oxygen sensors are located under this small aluminum housing Not illustrated in figure 2 see Figure 8 2 4 Power Supply The power supply circuit i...

Page 9: ...rs Upright plumb and level vertical orientation of the instrument is necessary for proper operation Mount the instrument on an appropriate vertical surface using the mounting flanges provided Avoid ar...

Page 10: ...play The instrument may go into alarm briefly but the sensors stabilize quickly If the instrument persists in alarm acknowledge the alarm by pressing the ALARM ACKN AUDIO DEFEAT switch If alarm persis...

Page 11: ...trip in the positions given in Table 3 Table 3 Outputs for 4 20mA for S N 599 and below Position Channel Function Range 16 CO Ground 4 mA 0 ppm 17 CO 4 to 20 mA 20 mA 100 ppm 18 DP Ground 4 mA 100 F 1...

Page 12: ...is activated If the oxygen content of the sample air decreases below the deficiency alarm set point the associated red LED is lit the associated relay chances state and the audio alarm is activated If...

Page 13: ...concentration is given in per cent by volume When sample flow is reduced below a limit the display switches from Flow yes to Flow no or from Flow y to Flow n 4 6 Operational Menu The operational menu...

Page 14: ...nt 4 7 2 Other Fault Indications Other fault indications are associated with sensor zero and calibration activities and are described in the maintenance section 5 0 of this manual 4 8 Dew Point Sensor...

Page 15: ...u is accessed with the OPTION and SELECT switches The maintenance menu diagram is shown in Figure 6 Maintenance Menu Flow Chart From the operational display press the OPTION switch twice run MAINTENAN...

Page 16: ...which locks the correct digit in place and moves the cursor one digit to the right Continue this process until the four digit key is complete When a valid key is inserted in this manner the display is...

Page 17: ...tandard and is best avoided The best zero gas is a cylinder of 20 9 oxygen in nitrogen with no CO present This is available in the calibration kit listed in section 6 0 Pressing the OPTION switch at t...

Page 18: ...ph 5 1 2 above When this procedure is complete ATTACH SPAN GAS SPAN 20 is displayed to indicate that the correct span gas for this procedure is 20 ppm CO in a mixture of 20 9 oxygen in nitrogen Do Not...

Page 19: ...rated with calibrating the CO sensor and the span voltage is close to but not greater then 0 38V an adjustment may be capable of being made to extend the life of the sensor Follow the procedure foe ca...

Page 20: ...T switch to lock the digit in place and move the cursor one digit to the right When all four digits of the new key have been selected NEW XXXX is displayed Record the new key without it the maintenanc...

Page 21: ...ur hours before recalibrating When recalibrating it is not necessary to make a voltage adjustment 5 3 Humidifier Tube The humidifier tube is used to assure that the CO and oxygen sensors are not subje...

Page 22: ...ove Shorting Clip Figure 9 Manifold Figure 10 Location of Parts for Replacement Manifold Cover Humidifier Tube Humidifier Tube Flow Control Orifice Oxygen Sensor CO Voltage adjustment Potentiometer CO...

Page 23: ...1000 Fuse 1 0 Amp 5x20mm 67016 1106 Sensor Oxygen 67020 1200 Sensor CO for units S N 299 and below 67016 1204 Sensor CO for units S N 300 and above 73070 009 Orifice 73108 002 Humidifier Tube 03401 00...

Page 24: ...ET shall not be liable for any loss or damage caused by the improper use of the product The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise thro...

Page 25: ...te the assessment of the degree of danger that he or she is exposed to Based upon knowledge of the effects of CO the Occupational Safety and Health Authority OSHA has set limits on exposure to CO in t...

Page 26: ...ENMET Corporation MED AIR 2000 22 Notes...

Page 27: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum 3...

Page 28: ......

Page 29: ...___________ Your Phone _______________________ Your PO Reference Number _______________ Your FAX _______________________ Payment Terms T COD Check one T VISA MasterCard______________________ ________...

Page 30: ......

Reviews: