Diagnostics and troubleshooting
Viomax CAS51D
30
Hauser
7.2
Cyclic cleaning
Compressed air is most suitable for automatic cyclic cleaning. There is a connection for
compressed air on every sensor. The cleaning unit, which is supplied with the device or can
be retrofitted, operates effectively at a rate of 20 l/min (5.4 US gal/min).
Type of soiling
Cleaning interval
Cleaning duration
Severe fouling with rapid buildup
5 min
10 s
Low risk of soiling
10 min
10 s
8
Diagnostics and troubleshooting
When troubleshooting, the entire measuring point must be taken into account:
• Transmitter
• Electrical connections and cables
• Assembly
• Sensor
The possible causes of error in the following table relate primarily to the sensor.
Problem
Testing
Solution
No display, no sensor
reaction
• Power supplied to transmitter?
• Sensor connected correctly?
• Medium flow present?
• Buildup on optical windows?
1.
Connect mains voltage
2.
Connect sensor correctly
3.
Ensure medium is flowing
4.
Clean sensor
Display value too high
or too low
• Buildup on optical windows?
• Gas bubbles present?
• Sensor calibrated?
1.
Cleaning
2.
Eliminate gas bubbles
3.
Calibration
4.
Check data record and modify if necessary
5.
Examine in workshop with test unit
Display value
fluctuating greatly
Gas bubbles present?
1.
Eliminate gas bubbles
2.
Check mounting location and select a
different mounting location if necessary.
Please observe the troubleshooting instructions provided in the Operating Instructions
of the transmitter. Check the transmitter if necessary.
Summary of Contents for Viomax CAS51D
Page 39: ......
Page 40: ...www addresses endress com ...