Viomax CAS51D
Installation
Hauser
23
SAC sensors with a gap width of 40 mm
Mount the compressed air cleaning unit prior to installing the sensor in the measuring
point, or remove the sensor from the medium. Clean the sensor if necessary.
1.
Screw the air diffuser from the accessory kit into the mounting holes behind the
sensor gap, and tighten by hand as far as it will go.
2.
Connect the compressed air supply provided by the customer. The hose piece with
hose coupling provided with the sensor can be used if desired.
5.6
Post-installation check
Put the sensor into operation only if you can answer "yes" to the following questions:
• Are the sensor and cable undamaged?
• Is the orientation correct?
• Is the sensor installed in an assembly and not freely suspended from the cable?